Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2020, Page 7 of 15  Not logged in ELOG logo
ID Date Author Subject Run Number DUT Beam Energy Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngledown Events Data
  21   Sun May 10 19:39:34 2020 Felix Ulrich-PurBeam Diamond+pCT6 500V19Diamond + DSSD-pCT662.4nonenone EFE SID2none 101031 

Diamond+pCT6
Dia HV = +/- 250V

ADC, FIR, ped, noise, cal

  22   Sun May 10 20:26:25 2020 Felix Ulrich-PurBeam+Diamond 500V20Diamond100.4nonenone EFE SID2none 103032 

Diamond
Dia HV = +/- 250V

ADC, FIR, ped, noise, cal

  23   Sun May 10 20:41:43 2020 Felix Ulrich-PurBeam+Diamond 500V21Diamond194.3nonenone EFE SID2none 150960 

Diamond
Dia HV = +/- 250V

ADC, FIR, ped, noise, cal

  24   Sun May 10 22:04:51 2020 Christian IrmlerBeam+Diamond 500V22Diamond252.7nonenone EFE SID2none 221012 

Diamond
Dia HV = +/- 250V

run taken instead of 800MeV --> beam parameters for 800 were not prepared

  25   Sun May 10 22:21:11 2020 Felix Ulrich-PurCalibration Tracker+Diamond 500V24Diamond + DSSD tracker252.7nonenone    none   

Diamond+Tracker
Dia HV = +/- 250V

 

  26   Fri Jun 12 23:10:40 2020 Felix Ulrich-PurCalibration TERA MPV 2001TERA-Cal800106020106dB none   

Setting 4

was stopped, cause timout setting was not adjusted to spill structure

  27   Fri Jun 12 23:47:07 2020 Felix Ulrich-PurCalibration TERA MPV 2002TERA-Cal800106020106dB none   

Setting 4

slice R1 did not converge therofor 9 steps

  28   Sat Jun 13 00:01:43 2020 Felix Ulrich-PurCalibration TERA MPV 3003TERA-Cal800106020106dB none   

Setting 4

 

  29   Sat Jun 13 00:14:44 2020 Felix Ulrich-PurCalibration TERA MPV 4004TERA-Cal800106020106dB none   

Setting 4

stopped because one slice could not be calibrated due to SW error

  30   Sat Jun 13 00:29:22 2020 Felix Ulrich-PurCalibration TERA MPV 4005TERA-Cal800106020106dB none   

Setting 4

S9 did not converge run stopped

 

  31   Sat Jun 13 00:34:32 2020 Felix Ulrich-PurCalibration TERA MPV 4006TERA-Cal800106020106dB none   

Setting 4

 

  32   Sat Jun 13 00:54:47 2020 Felix Ulrich-PurCalibration TERA MPV 3007TERA-Cal800106020106dB none   

Setting 4

repeated 300, because of SW error in previous run

 

  33   Sat Jun 13 01:07:52 2020 Felix Ulrich-PurCalibration TERA MPV 2008TERA-Cal800106020106dB none   

Setting 4

repeated 200, because of SW error in previous run

 

  34   Sat Jun 13 01:29:35 2020 Felix Ulrich-PurCalibration TERA MPV 6009TERA-Cal800106020106dB none   

Setting 4

few slices did not converge -> adapt range setting

 

  35   Sat Jun 13 01:51:17 2020 Felix Ulrich-PurCalibration TERA MPV 60010TERA-Cal800106020106dB none   

Setting 4

few slices did not converge -> adapted range setting

q9 and p3 did not converge-->choose mean voltage of group

 

  36   Sat Jun 13 02:16:06 2020 Felix Ulrich-PurCalibration TERA Sr-90 Voltages11TERA-Cal800106020106dB none 60000 

1Setting 4

used same voltages as calibrated with Sr-90 (MPV 300)

 

  37   Sat Jun 13 02:16:43 2020 Felix Ulrich-PurLinearity measurement TERA 100.4 gain 200MPV12TERA-Cal100.4BETA  EFE SID2none 200000 

Linearity Messung Gain 200MPV

  38   Sat Jun 13 02:37:10 2020 Felix Ulrich-PurLinearity measurement TERA 100.4 gain 300MPV13TERA-Cal100.4BETA  EFE SID2none 200000 

Linearity Messung Gain 300MPV

  39   Sat Jun 13 02:41:54 2020 Felix Ulrich-PurLinearity measurement TERA 100.4 gain 400MPV14TERA-Cal100.4BETA  EFE SID2none 200000 

Linearity Messung Gain 400MPV

  40   Sat Jun 13 03:06:25 2020 Felix Ulrich-PurLinearity measurement TERA 100.4 gain 600MPV15TERA-Cal100.4BETA  EFE SID2none 200000 

Linearity Messung Gain 600MPV error in config file aborted

ELOG V3.1.5-fc6679b