ID |
Date |
Author |
Subject |
Run Number |
DUT |
Beam Energy |
Beam Betatron |
Beam Chopper |
Beam Degrader |
Beam EFE |
Beam Orb |
Beam QuadScan |
PhantomType |
PhantomAngle |
Events |
Data |
24
|
Sun May 10 22:04:51 2020 |
Christian Irmler | Beam+Diamond 500V | 22 | Diamond | 252.7 | none | none | | EFE | | SID2 | none | | 221012 | |
Diamond
Dia HV = +/- 250V
run taken instead of 800MeV --> beam parameters for 800 were not prepared |
25
|
Sun May 10 22:21:11 2020 |
Felix Ulrich-Pur | Calibration Tracker+Diamond 500V | 24 | Diamond + DSSD tracker | 252.7 | none | none | | | | | none | | | |
Diamond+Tracker
Dia HV = +/- 250V
|
26
|
Fri Jun 12 23:10:40 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 200 | 1 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
was stopped, cause timout setting was not adjusted to spill structure |
27
|
Fri Jun 12 23:47:07 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 200 | 2 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
slice R1 did not converge therofor 9 steps |
28
|
Sat Jun 13 00:01:43 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 300 | 3 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
|
29
|
Sat Jun 13 00:14:44 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 400 | 4 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
stopped because one slice could not be calibrated due to SW error |
30
|
Sat Jun 13 00:29:22 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 400 | 5 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
S9 did not converge run stopped
|
31
|
Sat Jun 13 00:34:32 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 400 | 6 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
|
32
|
Sat Jun 13 00:54:47 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 300 | 7 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
repeated 300, because of SW error in previous run
|
33
|
Sat Jun 13 01:07:52 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 200 | 8 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
repeated 200, because of SW error in previous run
|
34
|
Sat Jun 13 01:29:35 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 600 | 9 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
few slices did not converge -> adapt range setting
|
35
|
Sat Jun 13 01:51:17 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 600 | 10 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
few slices did not converge -> adapted range setting
q9 and p3 did not converge-->choose mean voltage of group
|
36
|
Sat Jun 13 02:16:06 2020 |
Felix Ulrich-Pur | Calibration TERA Sr-90 Voltages | 11 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | 60000 | |
1Setting 4
used same voltages as calibrated with Sr-90 (MPV 300)
|
94
|
Sun Jun 21 06:12:08 2020 |
Felix Ulrich-Pur | Silimond 800MeV 500V | 32 | Diamond + DSSD tracker | 800 | 10 | 60 | 20 | 10 | 6dB | | | | 24116 | |
800 MeV global run (tracking effciency)
setting4 |
95
|
Sun Jun 21 07:53:56 2020 |
Felix Ulrich-Pur | Diamond only 800MeV 500V | 33 | Diamond | 800 | 10 | 60 | 20 | 10 | 6dB | | | | 109363 | |
800 MeV global run (tracking effciency)
setting4
telescope shifted 6mm to the left (in beam direction) |
96
|
Sun Jun 21 07:54:38 2020 |
Felix Ulrich-Pur | Diamond only 800MeV 600V | 34 | Diamond | 800 | 10 | 60 | 20 | 10 | 6dB | | | | 101831 | |
800 MeV global run (tracking effciency)
setting4
telescope shifted 6mm to the left (in beam direction) |
97
|
Sun Jun 21 08:15:13 2020 |
Felix Ulrich-Pur | Diamond only 800MeV 700V | 35 | Diamond | 800 | 10 | 60 | 20 | 10 | 6dB | | | | 101522 | |
800 MeV global run (tracking effciency)
setting4
telescope shifted 6mm to the left (in beam direction)
4 strips on n- side started to getted really noisy (408,409,483,484) |
Attachment 1: pct_dia_01_diamond_700V_beam_LocalHardware_02-000035_hitmap_n.png
|
 |
98
|
Sun Jun 21 08:43:23 2020 |
Felix Ulrich-Pur | Diamond only 800MeV 400V | 36 | Diamond | 800 | 10 | 60 | 20 | 10 | 6dB | | | | 101680 | |
800 MeV global run (tracking effciency)
setting4
telescope shifted 6mm to the left (in beam direction)
strips are not noisy |
99
|
Sun Jun 21 09:11:38 2020 |
Felix Ulrich-Pur | Diamond only 800MeV 300V | 37 | Diamond | 800 | 10 | 60 | 20 | 10 | 6dB | | | | 116365 | |
800 MeV global run (tracking effciency)
setting4
telescope shifted 6mm to the left (in beam direction)
|
100
|
Sun Jun 21 09:36:52 2020 |
Felix Ulrich-Pur | Silimond 800MeV 800V | 38 | Diamond + DSSD tracker | 800 | 10 | 60 | 20 | 10 | 6dB | | | | 33035 | |
800 MeV global run (tracking effciency)
setting4
telescope shifted 6mm to the left (in beam direction)
|
Attachment 1: pct_dia_01_silimond_800V_beam_LocalHardware_02-000038_hitmap.png
|
 |