ID Date Author Subject Run Number DUT Beam Energy Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orbdown Beam QuadScan PhantomType PhantomAngle Events Data
  26   Fri Jun 12 23:10:40 2020 Felix Ulrich-PurCalibration TERA MPV 2001TERA-Cal800106020106dB none   

Setting 4

was stopped, cause timout setting was not adjusted to spill structure

  27   Fri Jun 12 23:47:07 2020 Felix Ulrich-PurCalibration TERA MPV 2002TERA-Cal800106020106dB none   

Setting 4

slice R1 did not converge therofor 9 steps

  28   Sat Jun 13 00:01:43 2020 Felix Ulrich-PurCalibration TERA MPV 3003TERA-Cal800106020106dB none   

Setting 4

 

  29   Sat Jun 13 00:14:44 2020 Felix Ulrich-PurCalibration TERA MPV 4004TERA-Cal800106020106dB none   

Setting 4

stopped because one slice could not be calibrated due to SW error

  30   Sat Jun 13 00:29:22 2020 Felix Ulrich-PurCalibration TERA MPV 4005TERA-Cal800106020106dB none   

Setting 4

S9 did not converge run stopped

 

  31   Sat Jun 13 00:34:32 2020 Felix Ulrich-PurCalibration TERA MPV 4006TERA-Cal800106020106dB none   

Setting 4

 

  32   Sat Jun 13 00:54:47 2020 Felix Ulrich-PurCalibration TERA MPV 3007TERA-Cal800106020106dB none   

Setting 4

repeated 300, because of SW error in previous run

 

  33   Sat Jun 13 01:07:52 2020 Felix Ulrich-PurCalibration TERA MPV 2008TERA-Cal800106020106dB none   

Setting 4

repeated 200, because of SW error in previous run

 

  34   Sat Jun 13 01:29:35 2020 Felix Ulrich-PurCalibration TERA MPV 6009TERA-Cal800106020106dB none   

Setting 4

few slices did not converge -> adapt range setting

 

  35   Sat Jun 13 01:51:17 2020 Felix Ulrich-PurCalibration TERA MPV 60010TERA-Cal800106020106dB none   

Setting 4

few slices did not converge -> adapted range setting

q9 and p3 did not converge-->choose mean voltage of group

 

  36   Sat Jun 13 02:16:06 2020 Felix Ulrich-PurCalibration TERA Sr-90 Voltages11TERA-Cal800106020106dB none 60000 

1Setting 4

used same voltages as calibrated with Sr-90 (MPV 300)

 

  94   Sun Jun 21 06:12:08 2020 Felix Ulrich-PurSilimond 800MeV 500V32Diamond + DSSD tracker800106020106dB   24116 

800 MeV global run (tracking effciency)
setting4

  95   Sun Jun 21 07:53:56 2020 Felix Ulrich-PurDiamond only 800MeV 500V33Diamond800106020106dB   109363 

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

  96   Sun Jun 21 07:54:38 2020 Felix Ulrich-PurDiamond only 800MeV 600V34Diamond800106020106dB   101831 

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

  97   Sun Jun 21 08:15:13 2020 Felix Ulrich-PurDiamond only 800MeV 700V35Diamond800106020106dB   101522 

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 

4 strips on n- side started to getted really noisy (408,409,483,484)

  98   Sun Jun 21 08:43:23 2020 Felix Ulrich-PurDiamond only 800MeV 400V36Diamond800106020106dB   101680 

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

strips are not noisy

  99   Sun Jun 21 09:11:38 2020 Felix Ulrich-PurDiamond only 800MeV 300V37Diamond800106020106dB   116365 

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 

  100   Sun Jun 21 09:36:52 2020 Felix Ulrich-PurSilimond 800MeV 800V38Diamond + DSSD tracker800106020106dB   33035 

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 

  101   Sun Jun 21 10:21:53 2020 Felix Ulrich-PurSilimond 800MeV 800V39Diamond + DSSD tracker800106020106dB   2498 

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 

  102   Sun Jun 21 10:22:37 2020 Felix Ulrich-PurDiamond only 800MeV 800V40Diamond800106020106dB   108625 

800 MeV global run (tracking effciency)
setting4

 

telescope shifted 6mm to the left (in beam direction)

 