ID Date Author Subject Run Numberdown DUT Beam Energy Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle Events Data
  130   Sat Jul 4 04:43:54 2020 Felix Ulrich-PurBeta+EFE spot-size measurement20Lynx+RaMon+Octavius83yes  yes      

tested 180s exposure time of Lynx

  174   Sat Oct 10 11:53:27 2020 Manuel Christanell2020-10-1020TERA+Tracker100.4   1 SID2    

Start 11:45

  5   Sun May 10 15:58:42 2020 Christian Irmleralignment2Diamond + DSSD tracker252.7nonenone EFE SID2none 42740 

alignment run
DSSD HV = +/- 50V
Dia HV = +/- 250V
xml = pct_dia_01_silimond_beam_01-000001.xml

after ~ 1h beam (pumping)

start: 15:58
stop:  16:26

Attachment 1: 01-000002.root-cogLocations_5.1.1.0-Empirics.png
01-000002.root-cogLocations_5.1.1.0-Empirics.png
Attachment 2: 01-000002.root-cogLocations_5.1.1.1-Empirics.png
01-000002.root-cogLocations_5.1.1.1-Empirics.png
Attachment 3: 01-000002.root-cogLocations_5.1.2.0-Empirics.png
01-000002.root-cogLocations_5.1.2.0-Empirics.png
Attachment 4: 01-000002.root-cogLocations_5.1.2.1-Empirics.png
01-000002.root-cogLocations_5.1.2.1-Empirics.png
Attachment 5: 01-000002.root-cogLocations_5.1.3.0-Empirics.png
01-000002.root-cogLocations_5.1.3.0-Empirics.png
Attachment 6: 01-000002.root-cogLocations_5.1.3.1-Empirics.png
01-000002.root-cogLocations_5.1.3.1-Empirics.png
Attachment 7: 01-000002.root-cogLocations_5.1.4.0-Empirics.png
01-000002.root-cogLocations_5.1.4.0-Empirics.png
Attachment 8: 01-000002.root-cogLocations_5.1.4.1-Empirics.png
01-000002.root-cogLocations_5.1.4.1-Empirics.png
  27   Fri Jun 12 23:47:07 2020 Felix Ulrich-PurCalibration TERA MPV 2002TERA-Cal800106020106dB none   

Setting 4

slice R1 did not converge therofor 9 steps

  112   Sat Jul 4 03:59:45 2020 Felix Ulrich-PurEFE SID0 Energy measurement2Peakfinder145.4   yes SID0    

2 spills per measurement point

  141   Sun Sep 20 06:43:10 2020 Alexander Burker2020-09-202TERA-Cal800      none   

800 MeV, "Setting 4"

Timeout stop

  156   Sat Oct 10 09:38:12 2020 Alexander Burker2020-10-102TERA-Cal800          

Start 06:58; "Setting 4", Calibration

 

  279   Sun Dec 6 15:41:31 2020 Christian IrmlerMPW2199W13_04252.7   1 SID0 HV = 40V10 spills 

HV = 40V

r0 c2

  278   Sun Dec 6 15:38:59 2020 Christian IrmlerMPW2198W13_04252.7   1 SID0 HV = 60V10 spills 

HV = 60V

r0 c2

  277   Sun Dec 6 15:37:05 2020 Christian IrmlerMPW2197W13_04252.7   1 SID0 HV = 80V10 spills 

HV = 80V

r0 c2

  276   Sun Dec 6 13:59:35 2020 Christian IrmlerMPW2196W13_04252.7   1 SID0    

Testrun to check number of hits per splill --> ~5k hits per spill

  21   Sun May 10 19:39:34 2020 Felix Ulrich-PurBeam Diamond+pCT6 500V19Diamond + DSSD-pCT662.4nonenone EFE SID2none 101031 

Diamond+pCT6
Dia HV = +/- 250V

ADC, FIR, ped, noise, cal

Attachment 1: 01-000019.root-cogLocations_5.1.1.0-Empirics.png
01-000019.root-cogLocations_5.1.1.0-Empirics.png
Attachment 2: 01-000019.root-cogLocations_5.1.1.1-Empirics.png
01-000019.root-cogLocations_5.1.1.1-Empirics.png
  44   Sat Jun 13 03:08:39 2020 Felix Ulrich-PurLinearity measurement TERA 100.4 gain Sr9019TERA-Cal100.4BETA  EFE SID2none 200000 

9Linearity Messung Gain Sr-90

Calibration from Lab not useful-> maybe different voltages if all slices are connected

  129   Sat Jul 4 04:43:36 2020 Felix Ulrich-PurBeta+EFE spot-size measurement19Lynx+RaMon+Octavius100.4yes  yes      

tested 180s exposure time of Lynx

  173   Sat Oct 10 11:51:19 2020 Manuel Christanell2020-10-1019TERA+Tracker100.4   1 SID2  300.000 

Start 11:38

  20   Sun May 10 19:28:36 2020 Felix Ulrich-PurBeam Diamond 600V18Diamond62.4nonenone EFE SID2none 62250 

Diamond+pCT6
Dia HV = +/- 300V

ADC, FIR, ped, noise, cal

 

PARTII

  43   Sat Jun 13 03:08:25 2020 Felix Ulrich-PurLinearity measurement TERA 100.4 gain 400MPV18TERA-Cal100.4BETA  EFE SID2none 200000 

Linearity Messung Gain 400MPV previous voltages were not applied

  128   Sat Jul 4 04:16:41 2020 Felix Ulrich-PurBeta+EFE spot-size measurement18Lynx+RaMon+Octavius145.4yes  yes      

tested 180s exposure time of Lynx

  172   Sat Oct 10 11:49:36 2020 Manuel Christanell2020-10-1018TERA+Tracker100.41  1      

Start 11:30; with Tracker

  19   Sun May 10 19:27:52 2020 Felix Ulrich-PurBeam Diamond 600V17Diamond62.4nonenone EFE SID2none 44667 

Diamond+pCT6
Dia HV = +/- 300V

ADC, FIR, ped, noise, cal

 

unterbrochen, da TLU-ERROR