Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-MA2018 testbeam-MA2019 testbeam-MA2020 testbeam-MA2021 testbeam-MA2023 testbeam-MA2024
  MedAustron 2020, Page 11 of 15  Not logged in ELOG logo
ID Date Author Subject Run Number DUT Beam Energy Beam Betatron Beam Chopper Beam Degrader Beam EFE Beam Orb Beam QuadScan PhantomType PhantomAngle Eventsdown Data
  14   Sun May 10 17:44:50 2020 Felix Ulrich-PurCalibration Diamond+pCT6 500V11Diamond + DSSD-pCT6252.7nonenone    none   

Diamond+pCT6
Dia HV = +/- 250V

ADC, FIR, ped, noise, cal

  25   Sun May 10 22:21:11 2020 Felix Ulrich-PurCalibration Tracker+Diamond 500V24Diamond + DSSD tracker252.7nonenone    none   

Diamond+Tracker
Dia HV = +/- 250V

 

  26   Fri Jun 12 23:10:40 2020 Felix Ulrich-PurCalibration TERA MPV 2001TERA-Cal800106020106dB none   

Setting 4

was stopped, cause timout setting was not adjusted to spill structure

  27   Fri Jun 12 23:47:07 2020 Felix Ulrich-PurCalibration TERA MPV 2002TERA-Cal800106020106dB none   

Setting 4

slice R1 did not converge therofor 9 steps

  28   Sat Jun 13 00:01:43 2020 Felix Ulrich-PurCalibration TERA MPV 3003TERA-Cal800106020106dB none   

Setting 4

 

  29   Sat Jun 13 00:14:44 2020 Felix Ulrich-PurCalibration TERA MPV 4004TERA-Cal800106020106dB none   

Setting 4

stopped because one slice could not be calibrated due to SW error

  30   Sat Jun 13 00:29:22 2020 Felix Ulrich-PurCalibration TERA MPV 4005TERA-Cal800106020106dB none   

Setting 4

S9 did not converge run stopped

 

  31   Sat Jun 13 00:34:32 2020 Felix Ulrich-PurCalibration TERA MPV 4006TERA-Cal800106020106dB none   

Setting 4

 

  32   Sat Jun 13 00:54:47 2020 Felix Ulrich-PurCalibration TERA MPV 3007TERA-Cal800106020106dB none   

Setting 4

repeated 300, because of SW error in previous run

 

  33   Sat Jun 13 01:07:52 2020 Felix Ulrich-PurCalibration TERA MPV 2008TERA-Cal800106020106dB none   

Setting 4

repeated 200, because of SW error in previous run

 

  34   Sat Jun 13 01:29:35 2020 Felix Ulrich-PurCalibration TERA MPV 6009TERA-Cal800106020106dB none   

Setting 4

few slices did not converge -> adapt range setting

 

  35   Sat Jun 13 01:51:17 2020 Felix Ulrich-PurCalibration TERA MPV 60010TERA-Cal800106020106dB none   

Setting 4

few slices did not converge -> adapted range setting

q9 and p3 did not converge-->choose mean voltage of group

 

  92   Fri Jun 19 19:22:10 2020 Felix Ulrich-PurCalibration Diamond,Silimond,pCT6+Diamand30Diamond           

300V-800V in 100V steps

xml files:

/mnt/data/xmls/local/pct_dia_01_diamond_800V_02-000030.xml
/mnt/data/xmls/global/pct_dia_01_diamond_800V_beam_02-000030.xml

all calibration files

800 is only example replace with according voltage

  93   Sun Jun 21 05:31:37 2020 Felix Ulrich-PurSilimond31Diamond + DSSD trackerNoise          

local run

  107   Sun Jun 21 13:54:57 2020 Felix Ulrich-PurDiamond only 252.7 MeV 500V45Diamond252.7   EFE SID0    

252.7 MeV 4MHz flux text no spill structure  in current visible

 

telescope shifted 6mm to the left (in beam direction)

 

  108   Sun Jun 21 13:57:57 2020 Felix Ulrich-PurDiamond only 252.7 MeV 500V46Diamond252.7  10       

252.7 MeV 400MHz flux test: spill structure  in current visible

nominal flux degrader 10, random trigger

 

telescope shifted 6mm to the left (in beam direction)

 

  110   Sun Jun 21 14:17:34 2020 Felix Ulrich-PurSilimond 500V48DiamondNoise          

500 V run after high flux irradiation

 

  111   Sat Jul 4 03:58:14 2020 Felix Ulrich-PurEFE SID0 Energy measurement1Peakfinder100.4   yes SID0    

3 spills per measurement point

  112   Sat Jul 4 03:59:45 2020 Felix Ulrich-PurEFE SID0 Energy measurement2Peakfinder145.4   yes SID0    

2 spills per measurement point

  113   Sat Jul 4 04:00:05 2020 Felix Ulrich-PurEFE SID0 Energy measurement3Peakfinder62.4   yes SID0    

1 spills per measurement point

ELOG V3.1.5-fc6679b