ID |
Date |
Author |
Subject |
Run Number |
DUT |
Beam Energy |
Beam Betatron |
Beam Chopper |
Beam Degrader |
Beam EFE |
Beam Orb |
Beam QuadScan |
PhantomType |
PhantomAngle |
Events |
Data |
14
|
Sun May 10 17:44:50 2020 |
Felix Ulrich-Pur | Calibration Diamond+pCT6 500V | 11 | Diamond + DSSD-pCT6 | 252.7 | none | none | | | | | none | | | |
Diamond+pCT6
Dia HV = +/- 250V
ADC, FIR, ped, noise, cal |
25
|
Sun May 10 22:21:11 2020 |
Felix Ulrich-Pur | Calibration Tracker+Diamond 500V | 24 | Diamond + DSSD tracker | 252.7 | none | none | | | | | none | | | |
Diamond+Tracker
Dia HV = +/- 250V
|
26
|
Fri Jun 12 23:10:40 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 200 | 1 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
was stopped, cause timout setting was not adjusted to spill structure |
27
|
Fri Jun 12 23:47:07 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 200 | 2 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
slice R1 did not converge therofor 9 steps |
28
|
Sat Jun 13 00:01:43 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 300 | 3 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
|
29
|
Sat Jun 13 00:14:44 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 400 | 4 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
stopped because one slice could not be calibrated due to SW error |
30
|
Sat Jun 13 00:29:22 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 400 | 5 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
S9 did not converge run stopped
|
31
|
Sat Jun 13 00:34:32 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 400 | 6 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
|
32
|
Sat Jun 13 00:54:47 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 300 | 7 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
repeated 300, because of SW error in previous run
|
33
|
Sat Jun 13 01:07:52 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 200 | 8 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
repeated 200, because of SW error in previous run
|
34
|
Sat Jun 13 01:29:35 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 600 | 9 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
few slices did not converge -> adapt range setting
|
35
|
Sat Jun 13 01:51:17 2020 |
Felix Ulrich-Pur | Calibration TERA MPV 600 | 10 | TERA-Cal | 800 | 10 | 60 | 20 | 10 | 6dB | | none | | | |
Setting 4
few slices did not converge -> adapted range setting
q9 and p3 did not converge-->choose mean voltage of group
|
92
|
Fri Jun 19 19:22:10 2020 |
Felix Ulrich-Pur | Calibration Diamond,Silimond,pCT6+Diamand | 30 | Diamond | | | | | | | | | | | |
300V-800V in 100V steps
xml files:
/mnt/data/xmls/local/pct_dia_01_diamond_800V_02-000030.xml
/mnt/data/xmls/global/pct_dia_01_diamond_800V_beam_02-000030.xml
all calibration files
800 is only example replace with according voltage |
93
|
Sun Jun 21 05:31:37 2020 |
Felix Ulrich-Pur | Silimond | 31 | Diamond + DSSD tracker | Noise | | | | | | | | | | |
local run |
107
|
Sun Jun 21 13:54:57 2020 |
Felix Ulrich-Pur | Diamond only 252.7 MeV 500V | 45 | Diamond | 252.7 | | | | EFE | | SID0 | | | | |
252.7 MeV 4MHz flux text no spill structure in current visible
telescope shifted 6mm to the left (in beam direction)
|
108
|
Sun Jun 21 13:57:57 2020 |
Felix Ulrich-Pur | Diamond only 252.7 MeV 500V | 46 | Diamond | 252.7 | | | 10 | | | | | | | |
252.7 MeV 400MHz flux test: spill structure in current visible
nominal flux degrader 10, random trigger
telescope shifted 6mm to the left (in beam direction)
|
110
|
Sun Jun 21 14:17:34 2020 |
Felix Ulrich-Pur | Silimond 500V | 48 | Diamond | Noise | | | | | | | | | | |
500 V run after high flux irradiation
|
111
|
Sat Jul 4 03:58:14 2020 |
Felix Ulrich-Pur | EFE SID0 Energy measurement | 1 | Peakfinder | 100.4 | | | | yes | | SID0 | | | | |
3 spills per measurement point |
112
|
Sat Jul 4 03:59:45 2020 |
Felix Ulrich-Pur | EFE SID0 Energy measurement | 2 | Peakfinder | 145.4 | | | | yes | | SID0 | | | | |
2 spills per measurement point |
113
|
Sat Jul 4 04:00:05 2020 |
Felix Ulrich-Pur | EFE SID0 Energy measurement | 3 | Peakfinder | 62.4 | | | | yes | | SID0 | | | | |
1 spills per measurement point |