  HEPHY logbook of the Electronics Group, Page 3 of 4  Not logged in ELOG logo
Entry  Mon Nov 30 17:23:51 2015, Hao Yin, Belle II, system, HardwareRun001,  Room_bw_n.pngRoom_bw_p.png

ADC Hot, FIRRun001, PedestalRun001

room temp pedrun

target: BW

Entry  Mon Nov 30 17:47:54 2015, Hao Yin, Belle II, system, HardwareRun002,  Room_-z_n.pngRoom_-z_p.png

ADC Hot, FIRRun001, PedestalRun001

target: -Z

room temp pedrun

 

Entry  Mon Nov 30 18:32:13 2015, Hao Yin, Belle II, system, ADC Delay Scan Cold,  

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns)

Entry  Mon Nov 30 18:44:04 2015, Hao Yin, Belle II, system, FIRRun_Cold,  

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns)

 

Entry  Mon Nov 30 18:45:49 2015, Hao Yin, Belle II, system, PedestalRun_Cold,  

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

 

Entry  Mon Nov 30 18:49:23 2015, Hao Yin, Belle II, system, Hardware_Cold,  Cold_-z_n.pngCold_-z_p.png

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

target: -Z

 

Entry  Mon Nov 30 19:01:54 2015, Hao Yin, Belle II, system, Hardware_Cold,  Cold_ce_n.pngCold_ce_p.png

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

Temp = 0°C

target: CE

 

Entry  Mon Nov 30 19:48:05 2015, Hao Yin, Belle II, system, Hardware_Room_With_Cold_I2C,  RoomTemp_with_cold_config_ce_p.pngRoomTemp_with_cold_config_ce_n.png

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

with room temp i2c config

Temp = 0°C

target: CE

 

Entry  Mon Nov 30 20:06:15 2015, Hao Yin, Belle II, system, Hardware_Room_With_Cold_I2C,  RoomTemp_with_cold_config_-z_p.pngRoomTemp_with_cold_config_-z_n.png

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

with room temp i2c config

Temp = room

target: -Z

 

Entry  Mon Nov 30 20:16:37 2015, Hao Yin, Belle II, system, Hardware_Room_With_Cold_I2C,  RoomTemp_with_cold_config_-z_p_3.pngRoomTemp_with_cold_config_-z_n_3.png

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

ADC delay shifted by 3 (1.5ns) all sensors

with room temp i2c config

Temp = room

target: -Z

 

Entry  Mon Nov 30 20:27:42 2015, Hao Yin, Belle II, system, Hardware_Room_With_Cold_I2C,  RoomTemp_with_cold_config_-z_p_5.pngRoomTemp_with_cold_config_-z_n_5.png

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

ADC shifted by 5 (2.5ns) all sensors

with room temp i2c config

Temp = room

target: -Z

Note : n-side (strip nr 200 - 220 shadowhits)

Entry  Mon Nov 30 22:35:41 2015, Hao Yin, Belle II, system, Hardware_Room_With_Cold_I2C,  RoomTemp_with_cold_config_-z_p_9.pngRoomTemp_with_cold_config_-z_n_9.png

ADC Cold diff to hot: 1-3 adc delay config (max 1.5 ns), FIRRun_Cold_001

ADC shifted by 9 (4.5ns) all sensors

with room temp i2c config

Temp = room

target: -Z

Entry  Fri Apr 18 17:23:26 2008, Markus Friedl, SPS Testbeam June08, hybrid, hybrid 01, Noise of hybrid 01, sensor fully bonded, no HV hybrid01_sen_nohv_pednoise_apv0.pnghybrid01_sen_nohv_pednoise_apv1.png
HV bias not yet glued to backplane
Entry  Tue Apr 22 19:34:09 2008, Markus Friedl, SPS Testbeam June08, module, hybrid 01, Properties of hybrid 01, sensor fully bonded, HV=100V 6x
HV bias glued to backplane, HV=100V
Entry  Wed Apr 23 13:05:05 2008, Christian Irmler, SPS Testbeam June08, source, hybrid 01, time correlation between TDC and sensor measurement -> 3.1 ns RMS vie_run001_tpeak_vs_TDC.gif
 
Entry  Wed Apr 23 13:37:18 2008, Markus Friedl, SPS Testbeam June08, source, hybrid 01, analysis results of source test vie_run001_cluster_sig_0.gifvie_run001_cluster_noi_0.gifvie_run001_cluster_hit_0.gifvie_run001_cluster_clw_0.gif
Ignore the "KEK November 2007" title - that's a legacy and is already changed :-)

As of now, there is no distinction in 16 separate zones. However, the gaps between the the zones are clearly visible in the Hit Profile, as the edge strips on both sides have a larger sensitive area and thus collect more hits than other strips; hence the spikes in the (otherwise pretty gaussian) beam profile. There is a single strip with no entries in the center - that's the one that suffered from the bias bond repair action.
SNR=21 (peak mode) is pretty healthy and fits to similar detectors operated with APV25.
All data was taken in multi-peak mode with subsequent hit fitting to obtain amplitude and timing (see separate posting for timing precision).

The verbose output of the analysis is pasted below.


Analysis of vie_run001

Peak Mode, 3 x 200 initevents (first 10 skipped) + 99400 events
Number hybrids:  1	number zones:  2	 number sensors:  1
Using calibration file vie_cal001
No pedestal correction file
Seed/Neighbor/Cluster/Noisy Strips Cuts [RMS noise]: 5.0/3.0/5.0
Min. hitlength:   3


Comments:
SILC module 01
HV=100V, 40MHz, Tp=50ns, 30ns
Sr90 1mCi ,  black cloth cover



Analysis date: 23.04.2008 13:25:09


Analysis settings:
 runname: vie_run001_cluster
 clock: 40.00 MHz
 datafilepath: data/
 outputpath: output/
 subevents:  6
 fitmode: 2 (cal. fit)
 options: h




Results:

ModuleName           ZoneType    Ch    OKCh     OK%   Entries    MClW    MPSignal   Noise   MPSNR   HpSE  Occup
p_side            JP single sensor    256    256   100.0     90385    2.53    21546.1    729.4   20.68   0.53   1.81
Entry  Wed May 7 15:24:49 2008, Christian Irmler, SPS Testbeam June08, module, module 10/02, properties (noise, intcal), APVs bonded to the sensor 12x
Module tested with 1 and 2 rows bonded to the sensor, respectively. HV = 100 V Ibias (100V) = 19.1 nA Ibias (200V) = 23.7 nA
Entry  Wed May 7 15:26:24 2008, Christian Irmler, SPS Testbeam June08, module, module 20/09, properties (noise, intcal), APVs bonded to the sensor 12x
Module tested with 1 and 2 rows bonded to the sensor, respectively. HV = 100 V Ibias (100 V) = 25.1 nA Ibias (200 V) = 31.4 nA
Entry  Wed May 7 15:40:07 2008, Christian Irmler, SPS Testbeam June08, module, module 07/07, properties (noise, intcal), APVs bonded to the sensor 12x
Module tested with 1 and 2 rows bonded to the sensor, respectively. HV = 100 V Ibias (100 V) = 20.2 nA Ibias (200 V) = 21.9 nA
Entry  Wed May 7 16:12:58 2008, Christian Irmler, SPS Testbeam June08, module, module 12/08, properties (noise, intcal), APVs bonded to the sensor 12x
Module tested with 1 and 2 rows bonded to the sensor, respectively. HV = 100 V Ibias (100 V) = 22.2 nA Ibias (200 V) = 26.5 nA