Entry  Tue May 12 10:27:39 2015, Benedikt Würkner, SVD3 (H6B), Setup information, , , , , Unknown spider-config.jpg

Information concerning the SVD3 setup during a beam test

In the beam area setup the whole crate and the power supply with the usual spider cable and the two Keithly 2410 high voltage supplies. Monitoring of the 2410 happens via a NI MAX interface that connects to the GPIB-Ethernet interface and needs to be configured there. The IP and password are noted on the device. The software can be found on heros in medialib/LONGTERM/longterm R3.1_doubleSided as a normal *.vi file. 

Default adresses 16 and 17 as defined in labview, need to figure out on site which is which, if the voltage selected (e.g. 100V) is applied as +50/-50 or if it is applied as +100/-100 before connecting the cables. 

Take a run without dut as well to check the telescope integrity. Without dut in beamline!!!!!

Entry  Wed Jun 3 10:48:42 2015, Benedikt Würkner, common, Networking, , , , , Unknown 

WLAN Router:

SSID: Hephy

password: 20supersecret14

DHCP Range: 192.168.0.200 to 239

Netmask: /24

DNS Server: 192.168.0.254

Gateway: 192.168.0.254

external IP: 128.141.174.109 (für diverse Port Forwardings)

 

Manuell vergebene IP Adressen bitte unbedingt Benni mitteilen oder/und in der folgenden Tabelle eintragen! 

IP/MAC Zuweisung:

IP Machine/Name intern Port Forwarding (aussen)
192.168.0.124 SVD Kenwood  VNC(5900)  VNC(9000)
192.168.0.11 SVD3 DAQ  VNC(5900)  VNC(9001)
192.168.0.12 TUX DAQ  VNC(5900), VNC(5901), SSH(22)  VNC(9009), VNC(9002), SSH(9003),
192.168.0.13 Belle XY Tisch    
192.168.0.14 TUXOA  SSH(22), VNC(5900)  SSH(9004), VNC(9012)
192.168.0.20 JTAG Server    
192.168.0.148 alixcool    
192.168.0.16 slowcontrol  VNC(5900)  VNC(9005)
192.168.0.144 CAEN Power supply    
192.168.0.17 zenpixelwiki  SSH(22), VNC(5900)  SSH(9006), VNC(9008)
192.168.0.18 TLU (pcx25042, XY)  VNC(5900)  VNC(9007)
192.168.0.19 Cooling control PC VNC(5900) VNC(9011)
192.168.0.240 NI GPIB - ENET / 1000    
192.168.0.150 Printserver  VNC(5900)  VNC(9010)
       
137.138.113.254 Aida PC   vnc(5900)

 

Admin Zugang Wlan Router:

IP: 192.168.0.254

Username: admin

password: Hephy2014

 

Entry  Wed Jun 3 17:23:07 2015, Thomas Bergauer, Belle II FADC (H6A), XY stage positions, , , , , Unknown L5_Class_B-.PNG

x-Achse: positiv Richtung Saleve (counting room)

y-Achse: positiv Richtung UNTEN!!

 

position of laddertestbox xy stages:

Rev. Pos (BWD Sensor, BWD, Oben) x= 82.3, y=110.6

Mitte Sensor:

Pos 1: (mitte BWD sensor): x=144.74, y=80.8

Pos 2: (mitte Origami -z): x=269.62, y=80.8

Pos 3: (mitte Origami ce): x=394.42, y=80.8

Pos 4 (mitte FWD Micron Sensor): x=517, y=80.8

Ecke:

Pos 5 (Ecke BWD sensor): x=92.3, y=105.6

Pos 6 (Ecke Origami -z): x=217.18, y=105.6

Pos 7 (Ecke Origami ce): x=342.06, y=105.6

Pos 8 (Ecke FWD sensor): x=467,62, y=56,3 

Pos 9 (Ecke FWD sensor): x=568,33 y=64,48

Entry  Thu Jun 4 18:49:03 2015, Benedikt Würkner, SVD3 (H6B), Setup, , , , , Unknown IMG_20150604_184442.jpgIMG_20150604_184451.jpgIMG_20150604_184516.jpg
 
Entry  Sun Jun 7 14:44:29 2015, Thomas Bergauer, SVD3 (H6B), XY Stage and DUT Positions, , , , , Unknown FWD-BWD.PNGapv-numbering.jpgFWD-BWD_Bemaßung.PNG

Coordinate system:

pos x-Direction = Jura side

pos y-Direction = Down

Zero-point is in the top right corner in Beam direction

 

absolute Position of PI Stages:

Pos 1: x=100 mm, y=79 mm (Hits in APV 3 und 8-9 in p-Seite und 14 und 17 in n-Seite )

Pos 2: x=69.3 mm, y=79 mm (Hits in APV 3-4 und 8-9 in p-Seite und 13 und 18 in n-Seite) 

Pos 3: x=38.6 mm, y=84 mm

Pos 4: x=130.7 mm, y=84 mm

Pos 5: x=30.5 mm, y= 99.74 mm (wide side top)

Pos 6: x=136.08 mm, y= 91.49 mm (narrow side top)

Pos 7: x= 30.5 mm, y= 79 mm (wide side center)

Pos 8: x=136.08 mm, y= 79 mm (narrow side center)

Pos 9: x=30.5 mm, y= 60.67 mm (wide side down)

Pos 10: x= 136.08mm, y=68.9 mm (narrow side down)

Entry  Sun Jun 7 20:02:40 2015, Hao Yin, Belle II FADC (H6A), Cooling Box Emergency break, , , , , Unknown 

Something triggered a emergency break. HWRun012 only last 1-10k event has to be discarded.

Entry  Mon Jun 8 11:33:24 2015, Thomas Bergauer, common, beam profile, , , , , Unknown Beamprofile.png

120GeV hadrons, beam config H6A.BELLE.003

Entry  Mon Jun 8 16:53:13 2015, Benedikt Würkner, SVD3 (H6B), EUTelescope Fitter not working investigation, , , , , Unknown nice-alignment.pngbad-alignment.png

Apparently every problem can be attributed to a misalignment problem. The telescope has troubles with heavily misaligned DUTs. Was resolved by creating different gear-files for each run with pretty precise coordinates of the DUT position calculated for the telescope coordinate system. 

Additionally one has to watch out for misaligned planes of the telescope itself. It was required to allow for more than 2mm offset for the telescope planes behind the DUTs. This can be easily checked by looking at the PreAligner plots in the -hitmaker-histo.root file.

It look something like this:

and not like this:

 

Run number offset value Comment
49 -1  
50 -1 Expected default value. EORE=49401
51 -1 Expected default value. EORE=49400
52 -1 Checked -1 EORE= 49401
53 -1 Checked -1 EORE= 49401
54 -1 Checked -1 EORE= 119401, 
55 -1 Checked -1 EORE= 409388
56 -1 Checked -1 EORE= 49392
57 -1 Checked -1 EORE= 49377
58 -1 Checked -1 EORE= 49397

 

Run Position x-Table x-telescope FW x-telescope BW
49 1 100 -13,7 -16,7
50 2 69,3 17 14
51 3 38,6 47,7 44,7
52 4 130,7 -44,4 -47,4
53 5 30,5 55,8 52,8
54 6 136,8 -50,5 -53,5
55 7 30,5 55,8 52,8
56 8 136,8 -50,5 -53,5
57 9 30,5 55,8 52,8
58 10 136,8 -50,5 -53,5
Entry  Mon Jun 8 18:08:49 2015, Thomas Bergauer, SVD3 (H6B), Telescope online monitor, , , , , Unknown 

a working version of the EUDET online monitor (which works also with the FE-I4) is at the following location

~/BELLETEST/eudaq-neu/build/monitors/onlinemon$ ./OnlineMon.exe -sc 0 -f /home/telescope/eudaq/data/run000060.raw

Entry  Tue Jun 9 10:01:03 2015, Lukas Bulla, SVD3 (H6B), TLU setup info, , , , , Unknown upstream.jpgdownstream.jpg

Channels:

  • Upstream vertical -> channel 0; (bitmask 1)
  • Upstream horizontal -> channel 1; (bitmask 2)
  • Downstream vertical -> channel 2; (bitmask 4)
  • Downstream horizontal ->channel 3; (bitmask 8)

 

TLU DuT interface:

DUT Interface Bitmask DAQ
1 1 Mimosa (EUDET Telescope)
2 2 leer
3 4 SVD3
4 8 FE-I4

 

Entry  Wed Jun 10 18:05:03 2015, Benedikt Würkner, SVD3 (H6B), FEI4 configuration, , , , No, Unknown 

There are two planes, one fixed upstream and one fixed downstream. 

CIS3WII_6 upstream and CIS2WI_IBL2 downstream.

They are configured using the pc in H6B which can be accessed via microsoft remote desktop from the network 192.168.5.x using the ip 192.168.5.6. 

The configuration files can be found in the folder "E:\ConfigurationCMS\" on the server and are named after the modules. 

Only one of the FEI4 detectors can be used at a time since they are of different generations (FEI4A and FEI4B) which can't be used at the same time with the available USB board. vnc

The configuration for EUDAQ needs to be adapted for the used module by changing the name of the configuration file. 

[Producer.USBpixI4]
SRAM_READOUT_AT = 5
SkipConfiguration = no
UseSingleBoardConfig = no
boards = 230
modules[230] = 1
#upstream
config_file = E:\ConfigurationCMS\CIS3_W11_06_thr1600_7to21ke_modified.cfg.root
#downstream:
#config_file = E:\ConfigurationCMS\CIS2_W02_IBL1_thr1600_7to14_modified.cfg.root
fpga_firmware = E:\icwiki_svn\USBPixI4\host\tags\release-5.3\config\usbpixi4.bit
lvl1_delay = 16
tlu_trigger_data_delay = 10

In my experience switching between the planes also requires a restart of the USBPix DAQ because apparently something is not working with the reload of a different config file afterwards. I had the phenomenon that the DAQ responded like it was working but the datastream was empty. Therefore always check the online monitor for hitmaps. 

Entry  Thu Jun 4 14:15:50 2015, Hao Yin, Belle II FADC (H6A), FIR, 001, , , , Good 

FIR filer and ADC delay with all sensor and telescope.
telescope n-side switched.

Entry  Thu Jun 4 14:16:20 2015, Hao Yin, Belle II FADC (H6A), Pedestal, 001, 10000, , , Good 

FirRun001, room temp
telescope n-side switched.

Entry  Thu Jun 4 15:12:41 2015, Hao Yin, Belle II FADC (H6A), Hardware, 001, 50000, 1, No, Good 20x

PedestalRun001, FIRRun001, room temp
telescope n-side switched.
In order to correct the position change the

Entry  Fri Jun 5 12:02:29 2015, Hao Yin, Belle II FADC (H6A), Sixlet Calibration, 001, , , , Good 

room temp max subevent at 2. with lat = 96

Entry  Thu Jun 4 18:44:50 2015, Hao Yin, Belle II FADC (H6A), Hardware, 002, 50000, 2, No, Good 19x

room temp, FIRRun001, pedestal run 003

Entry  Fri Jun 5 12:03:34 2015, Hao Yin, Belle II FADC (H6A), Sixlet Calibration, 002, , , , Good 

lat = 97 to a greater range of lat scan

Entry  Sun Jun 7 10:39:56 2015, Hao Yin, Belle II FADC (H6A), Calibration, 002, , , , Good 

Cooled!
FIRRun002, pedestalrun006; temp see attachment in pedestalrun006
 

Entry  Thu Jun 4 18:45:42 2015, Hao Yin, Belle II FADC (H6A), Pedestal, 003, 10000, , , Good 

firrun003, room temp
 

Entry  Thu Jun 4 22:07:16 2015, Hao Yin, Belle II FADC (H6A), Hardware, 003, 189k, 3, No, Good 21x

CE. pedestal run 003. Fir run 002. Room temp