Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  Belle/Infineon/AIDA testbeam at SPS in Nov 2014, Page 2 of 6  Not logged in ELOG logo
ID Date Author Project Subjectup Run Number Events StartTime EndTime Data
  14   Mon Nov 17 22:21:41 2014 Hao Yin FIR run001 Mon Nov 17 22:08:29 CET 2014 Good

Excellent, brilliant run. 

  54   Thu Nov 20 10:33:13 2014 Benedikt WürknerBelle SVDFIRRun and Delay Scan003   Unknown

 Cooled see appendix

with new layout (l3_n-side channels changed and reactivated).

Latest FIRRun

Attachment 1: Temp_19-11-14.PNG
Temp_19-11-14.PNG
  23   Tue Nov 18 09:45:59 2014 Hao Yin FIRRun_002002 Tue Nov 18 09:45:58 CET 2014 Unknown

New FIR run after reconecting to original setting. 

  50   Thu Nov 20 00:52:50 2014 Hao YinBelle SVDFIRRun_002    Unknown

cooled with new delay run

  80   Fri Nov 21 19:05:27 2014 Hao YinBelle SVDFIRRun_004  Fri Nov 21 19:05:23 CET 2014 Unknown

New ADC delay scan. cooled, -25 temperature

 

  60   Thu Nov 20 15:32:33 2014 Hao Yin HardwareRun010   Crap

same as hardware run 009

FIFO FULL ERROR starting at ~event 5000

  63   Thu Nov 20 16:13:18 2014 Hao YinBelle SVDHardwareRun011   Crap

cooled see appendix (FirRun003)

PedestalRun_009

multi 6 Raw Mode

FIFO size increased

 

 

FIFO FULL at  event ~27000

  66   Thu Nov 20 22:11:06 2014 Hao YinBelle SVDHardwareRun012   Good

cooled see appendix (FirRun003)

PedestalRun_009

multi 6 Raw Mode

FIFO size increased

  71   Fri Nov 21 15:02:28 2014 Benedikt WürknerBelle SVDHardwareRun0131433  Unknown

cooled see appendix (FirRun003)
PedestalRun_010
multi 6 Raw Mode
FIFO size increased
L3n deactivated
100000 deadtime (CTRL TRG)
Zero_suppressed

  72   Fri Nov 21 15:03:08 2014 Benedikt WürknerBelle SVDHardwareRun014989  Unknown

cooled see appendix (FirRun003)
PedestalRun_010
multi 6 Raw Mode
FIFO size increased
L3n deactivated
100000 deadtime (CTRL TRG)
Zero_suppressed

  73   Fri Nov 21 15:07:24 2014 Benedikt WürknerBelle SVDHardwareRun0191509482014-11-21 02:41:552014-11-21 10:11:02Unknown

cooled see appendix (FirRun003)
PedestalRun_010
multi 6 Raw Mode
FIFO size increased
L3n deactivated
100000 deadtime (CTRL TRG)

VME Error, overnight run

  74   Fri Nov 21 15:10:19 2014 Benedikt Würkner HardwareRun15-22 (exept 19) none  Crap
  93   Sat Nov 22 11:37:16 2014 Benedikt WürknerBelle SVDHardwareRun0364503782014-11-22 1:00:002014-11-22 10:27:54Unknown

 Not completely cooled since cooling failed at 1:42 in the night

Bias voltage set to +-60V

Some error with event no17 or 18 leading to the data not being easily analysable requiring a hack in TuxOA

Pedestal_013 (cooled)

Noisefile: HardwareRun_036.noi

Attachment 1: SPS_Testbeam-Temp.xlsx
Attachment 2: Screen_Shot_2014-11-22_at_12.41.44.png
Screen_Shot_2014-11-22_at_12.41.44.png
  94   Sat Nov 22 15:31:49 2014 Hao YinBelle SVDHardwareRun038437882014-11-22 15:18:40 Unknown

Cooled down to -20

Bias voltage set to +-60V

Pedestal_013 (cooled)

Noisefile: HardwareRun_036.noi

Zero suppressed

  105   Sun Nov 23 01:05:43 2014 Hao YinBelle SVDHardwareRun039200002014-11-22 15:18:40 Crap

Cooled down to -20

Bias voltage set to +-80V

Pedestal_013 (cooled)

Noisefile: HardwareRun_036.noi

 

  106   Sun Nov 23 01:37:42 2014 Hao YinBelle SVDHardwareRun040200002014-11-23 01:37:00 Unknown

Bias voltage set to +-80V

Pedestal_015 (cooled)

"New" layout with no cooling (CO2 is no more)

Newly set values for APC Delay

New FIR Run 005

Temperature around 30 C

Bias: +-80V

  111   Sun Nov 23 10:51:49 2014 Hao YinBelle SVDHardwareRun04276863  Unknown

Hardware Run +-40V Bias

Room Temperature

Pedestal 016

Zero Suppressed

Attachment 1: Temp_23-11-14.PNG
Temp_23-11-14.PNG
  112   Sun Nov 23 12:17:59 2014 Benedikt WürknerBelle SVDHardwareRun04420000  Unknown

Hardware Run +-30V Bias

Room Temperature

Pedestal 017 (again marked all the bad strips)

Zero Suppressed

  114   Sun Nov 23 12:55:47 2014 Benedikt WürknerBelle SVDHardwareRun04520000  Unknown

Hardware Run +-50V Bias

Room Temperature

Pedestal 018 (again marked all the bad strips)

Zero Suppressed

  118   Sun Nov 23 15:13:37 2014 Markus FriedlBelle SVDHardwareRun00310060023.11.2014 14:07:2323.11.2014 16:51:27Good

Hardware run, hitting APVs 0/1 on p-sides, 1/2 on n-sides

EUDAT telescope running with difference of 2 events

ELOG V3.1.5-fc6679b