  46   Fri Oct 7 18:40:07 2011 Christian IrmlerBELLEOrigami stack cooledcal0098380007.10.2011 18:35:04 07.10.2011 19:31:51Good

Origami6
HV=+-40V,
CO2 cooling: P_TT_4: 14.9 bar, mf=0.9 g/s
beam on; xy table position x=306 y=76.5

I=18.6µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages 

CO2 pressure reduced to 14.9 bar --> temp is about 4K colder
Temp. at pipe inlet (K7) = -18°C
Temp. at pipe outlet (K8) = -21.8°C

used multi6 config instead of single hit config (cern11_origami_cooled2_multi6.cfg)

currents:
07.10.2011  18:36:07  event=1581  rate=31events/s temp/curr=33.9/-05.3 33.9/-05.6 33.9/-06.6 32.8/001.4 32.8/001.9 32.8/018.1

  47   Fri Oct 7 19:35:14 2011 Christian IrmlerBELLEOrigami stack cooledrun01490k07.10.2011 19:49:2607.10.2011 23:26:52Good

Origami6
HV=+-40V,
CO2 cooling: P_TT_4: 14.9 bar, mf=0.9 g/s
beam on; xy table position x=306 y=74.5

centered

I=17.8µA; old I2C settings; corrected trigger delay (htr); adjusted front-end voltages 

CO2 pressure reduced to 14.9 bar --> temp is about 4K colder
Temp. at pipe inlet (K7) = -18°C
Temp. at pipe outlet (K8) = -21.8°C

currents:

07.10.2011  23:26:21  event=89786  rate=0events/s temp/curr=33.0/-05.3 33.0/-05.8 33.0/-06.6 31.9/001.4 32.0/001.9 31.9/017.7
 

  48   Sat Oct 8 00:14:37 2011 Christian IrmlerBELLEOrigami stackrun01518011108.10.2011 00:06:4108.10.2011 08:24:38Good

Origami6
HV=+-40V,
no cooling
beam on; xy table position x=306 y=74.5
Istart=57µA;
std. I2C settings

centered

currents:
8.10.2011  00:07:43  event=991  rate=0events/s temp/curr=33.0/-15.4 33.0/-17.3 33.0/-17.7 31.9/011.5 31.9/013.2 32.0/040.20
...

08.10.2011  06:32:34  event=179710  rate=0events/s temp/curr=33.1/-22.1 33.0/-25.6 33.1/-22.6 31.9/018.2 31.8/021.7 31.9/048.5
08.10.2011  06:33:28  event=180105  ***WARNING*** Error detected in APV 0, subevent 2. Event discarded, reset sent.
 

Permanent APV errors at several (all p-side, some n-side) APV chips after 180105 events, which could not be recovered by software resets. After restart of run, all APVs where alive again, hence hardware reset and I2C reconfiguration after 5 or 10 consecutive APV errors should be added to TUXDAQ software.

After APV errors occured, the cummulative bias current decrised from 78µA to 54µA. This effect can occures whenever the digital part of the APV is switched between idle and full operation. It seems that this is a thermal effect, caused by  the slightly higher power consumption of the digital part of the chips.

 

  55   Sun Oct 9 10:05:34 2011 Markus FriedlHPKbaby_stdcal0108380008.10.2011 13:36:2508.10.2011 13:54:07Good

HV=+-40V,
no cooling
beam off; xy table position x=  y=
HV= Istart=uA;
std. I2C settings
 

  52   Sat Oct 8 16:45:14 2011 Thomas BergauerHPKbaby_stdrun01815000008.10.2011 16:44:1308.10.2011 19:33:33Good

no cooling
beam on; xy table position x=322  y=80
HV=200V Istart_PY=8.2uA;  Istart_N=0.45uA
std. I2C settings, now with correct timing

  53   Sat Oct 8 20:33:58 2011 Karl-Heinz HoffmannHPKbaby_stdrun01915000008.10.2011 20:22:4508.10.2011 23:19:23Good

no cooling
beam on; xy table position x=333  y=80
HV=200V Istart_PY=8.2uA;  Istart_N=0.45uA
std. I2C settings

  54   Sat Oct 8 23:27:44 2011 Karl-Heinz HoffmannHPKbaby_stdrun02038200008.10.2011 23:24:1609.10.2011 09:24:20Good

no cooling
beam on; xy table position x=327,5  y=65
HV=200V Istart_PY=8.2uA;  Istart_N=0.45uA
std. I2C settings

  57   Sun Oct 9 11:00:33 2011 Karl-Heinz HoffmannHPKbaby_addrun02115000009.10.2011 10:56:4709.10.2011 12:47:23Good

no cooling
beam on; xy table position x=330  y=74
HV=200V Istart_PY=1,9uA;  Istart_N=0.27uA
std. I2C settings

  60   Sun Oct 9 13:12:32 2011 Karl-Heinz HoffmannHPKbaby_addrun022670k09.10.2011 13:09:3210.10.2011 00:48:16Good

no cooling
beam on; xy table position x=330  y=64
HV=200V Istart_PY=1,9uA;  Istart_N=0.27uA
std. I2C settings

  62   Mon Oct 10 10:37:52 2011 Christian IrmlerCNMCNM stackcal0128380010.10.2011 11:29:5510.10.2011 11:40:55Unknown

no cooling
beam on; xy table position x=330  y=64
HV_CNM=20V, HV_BSTD1=200V
Istart_CNM= 28 / 1 uA;  Istart_BSTD= 0.12 uA

  63   Mon Oct 10 12:15:52 2011 Manfred ValentanCNMCNM stackrun02411000010.10.2011 12:03:5810.10.2011 14:05:56Good

no cooling
beam on; xy table position x=320.5  y=79.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 28 / 1 uA;  Istart_BSTD= 0.12 uA

  66   Mon Oct 10 14:11:53 2011 Christian IrmlerCNMCNM stackrun02511000010.10.2011 14:09:0010.10.2011 16:11:57Good

no cooling
beam on; xy table position x=335.5  y=79.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 31 / 1.4 uA;  Istart_BSTD= 0.2 uA

  67   Mon Oct 10 16:26:20 2011 Marko DragicevicCNMCNM stackrun02630000010.10.2011 16:23:4610.10.2011 20:30:00Good

no cooling
beam on; xy table position x=328  y=69.5
HV_CNM=20V, HV_BSTD1-N=200V
Istart_CNM= 27 / 1.5 uA;  Istart_BSTD= 0.2 uA

  61   Mon Oct 10 00:50:23 2011 Christian IrmlerHPKbaby_addrun023680k10.10.2011 00:51:2610.10.2011 09:15:35Good

no cooling
beam on; xy table position x=330  y=74
HV=200V Istart_PY=1,9uA;  Istart_N=0.27uA
std. I2C settings

same as run021, but taking more events

  68   Mon Oct 10 21:54:24 2011 Marko DragicevicMicronMicron-Baby-Stack p-spray, run bottom rightrun027700k10.10.2011 22:40:3111.10.2011 09:49:11Unknown

hit modules in bottom left corner -> Zones narrow and half-narrow for COMB_IRRAD, Zones wide and half-wide for COMM_IRRAD
timing corrected w.r.t. run004

Position of HEPHY x-z-table:
x = 322
z = 70

Note: z-Position not identical with position of run004, x-z-table seemed to have moved vertically. This configuration now hits the same sensor position.

  72   Mon Feb 20 17:17:27 2012 Christian IrmlerBELLEOrigami Module #4run10050k20.02.2012 17:07:2120.02.2012 17:17:22Good

Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C

standard I²C settings for uncooled APV chips.

Lab measurement with new dark box but old, large scintillator.

20.02.2012  17:08:27  event=5771  ***WARNING*** Error detected in APV 6, subevent 0. Event discarded, reset sent. --> APV 6 recovered afterwards

  73   Mon Feb 20 17:20:20 2012 Christian IrmlerBELLEOrigami Module #4cal10050k20.02.2012 17:19:0320.02.2012 17:29:58Good

Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C

standard I²C settings for uncooled APV chips.

Lab measurement with new dark box but old, large scintillator.

Calibration run for run100.

  74   Mon Feb 20 17:27:02 2012 Christian IrmlerBELLEOrigami Module #4run1012449720.02.2012 17:33:3320.02.2012 17:38:21Good

Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=2,03µA
CO2 cooling on, Tin=-16°C, Tout=-16,7°C

I²C settings for cooled APV chips --> same as used in run010.

Lab measurement with new dark box but old, large scintillator.

CO2 empty after 24497 events --> run stopped

  75   Wed Feb 22 12:15:11 2012 Christian IrmlerBELLEOrigami Module #4run102170k22.02.2012 12:31:0822.02.2012 13:03:27 Good

same as run101

Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=1,8µA
CO2 cooling on, Tin=-20.2°C, Tout=-21.3°C

I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_multi6_cooled.cfg

Lab measurement with new dark box but old, large scintillator.

CO2 empty after 24497 events --> run stopped --> no corresponding calibration available, but for rought analysis cal100 can be used for this run, too

  76   Wed Feb 22 13:17:34 2012 Christian IrmlerBELLEOrigami Module #4cal1028380022.02.2012 13:11:3022.02.2012 13:25:18Good

Sr90, large scintillator, HVret @ Origami open
Origami Mod. #4
HV=+-40V, Ibias=1,8µA
CO2 cooling on, Tin=-20.2°C, Tout=-21.3°C

I²C settings for cooled APV chips --> same as used in run010
e:\cvi-projects\apvdaq_fadc\origami6_V2_single_cooled.cfg

Lab measurement with new dark box but old, large scintillator.

ELOG V3.1.5-fc6679b