93
|
Wed Oct 6 16:33:55 2010 |
Markus Friedl | Micron Baby | run | run051 | 90000 | 06.10.2010 16:32:03 | 06.10.2010 20:59:58 |
8 x Micron Baby (3 irradiated), no cooling
HV=100V I=209uA (start of run, still rising); 203uA (end of run)
beam on, NIM trigger from TLU (no handshake)
telescope position 131 / 53.5 / 7.5
no telescope run - being tuned (some intervals during data taking are possible)
|
96
|
Wed Oct 6 23:48:49 2010 |
Markus Friedl | BELLE | run | run052 | 165789 | 06.10.2010 23:47:12 | 07.10.2010 08:43:42 |
Belle stack without Origami2
HV=80V I=217uA
telescope setting 103 / 30 / 277.5
telescope run 20184
temperature of chiller set to 15°C
beam stopped at 7.10. 08:00 sharp, no beam until 8.10 - 8:00 according to SPS Page 1. We keep DAQ and telescope running and cross our fingers for at least some spills of particles.
Irradiated micron module 2, n-side: excessive hit count on channels at the edges of the APV chips - no such thing on the p-side |
97
|
Fri Oct 8 17:26:36 2010 |
Markus Friedl | BELLE | cal | cal015 | 83800 | 08.10.2010 17:22:44 | |
Belle stack without Origami2
HV=11V I=32uA
beam on
2 suspicious APVs on n side of Micron due to burnt strip
trying to identify problematic Micron n side channels with low bias voltage (APV22 baseline saturates at higher bias) |
99
|
Fri Oct 8 18:12:43 2010 |
Markus Friedl | BELLE | run | run054 | 90000 | 08.10.2010 18:08:48 | 08.10.2010 21:49:36 |
Belle stack w/o Origami2 after irradiation
HV=80V I=224uA (beginning of run)
2 suspicious APVs on Micron n side
xy-table position 7 / 30 / 277.5 (shoots at the smaller edge of the trapezoidal sensor)
telescope run 20195 |
100
|
Fri Oct 8 23:05:03 2010 |
Markus Friedl | BELLE | cal | cal016 | 83800 | 08.10.2010 23:03:16 | |
Belle stack w/o Origami2 after irradiation
HV=80V / 258 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron due to burnt strip
Bias connection of Origami repaired --> lower noise on p-side |
103
|
Sat Oct 9 00:17:02 2010 |
Markus Friedl | BELLE | run | run056 | 225k | 09.10.2010 00:15:53 | 09.10.2010 09:54:28 |
Belle stack w/o Origami2 after irradiation
HV=80V / I_start=255 µA, I_stop=243 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron (possibly due to burnt strip)
same position as in previous run
telescope run 20198 - now with correct telescope threshold settings
a lot of "Error detected in APV 21, subevent 2. Event discarded, reset sent." errors started at 09.10.2010 09:00:25 (after whole night without any error) |
106
|
Sat Oct 9 15:29:09 2010 |
Markus Friedl | BELLE | run | run058 | 30000 | 09.10.2010 15:26:50 | 09.10.2010 16:21:37 |
Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73 - now with multi6 (previous run was single)
HV=100V / 35uA (beginning of run) 34uA (end of run)
cooling: chiller set to 15C
telescope run 20201
Origami p-side does not receive triggers -> disabled
origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...
|
|
108
|
Sat Oct 9 19:44:36 2010 |
Markus Friedl | common | end of success | | | 09.10.2010 19:44:23 | NaN |
|
13
|
Mon Sep 27 15:39:33 2010 |
Thomas Bergauer | common | PI-XY Table | | | | |
M-521.dd: (links/rechts)
downstream rechts => pos. limit [0]
downstream links => neg. limit [207]
----------------------
M-511.dd: (oben/unten)
oben [102]
unten [0]
----------------------
M-060.dg: (rotation)
parallel beam [7,5]
-------------------------------------------
Remote Control:
lokal ssh tunnel starten:
ssh -g -C -L5959:zenpixelw:5900 eudet@eudetmac001.cern.ch
danach mit "Chicken of the VNC" (von lokalem Mac) VNC auf localhost (port 5959).
|
14
|
Mon Sep 27 15:41:28 2010 |
Thomas Bergauer | common | Szintillatoren | | | | |
1: breit (1)
2: schmal (2)
3: breit (4)
4: schmal (8)
AND aus beiden breiten: 5
AND aus beiden schmalen: 10
|
15
|
Mon Sep 27 15:42:09 2010 |
Thomas Bergauer | common | TLU: wir stecken auf DUT 3 | | | | |
TLUControl.exe ist auf zenpixell3 unter /home/eudet/eudaq/bin
TLUControl.exe -d 14 -a 12
-dXX Mask for DUT:
1 abgeschnittenes Dummykabel
2 EUDRB hinten
4 EUDRB vorne
8 DUT3(wir)
-t100 internal trigger
-aXX Mask for Szintis:
1 4 mm (front)
2 4 mm (front)
4 9 mm (back)
8 9 mm (back) |
16
|
Mon Sep 27 18:54:47 2010 |
Thomas Bergauer | BELLE | XY Position | | | | |
center position: 55/40/277,5 |
24
|
Tue Sep 28 16:50:31 2010 |
Thomas Bergauer | common | serial connection chiller | | | | |
Der chiller braucht ein gekreuztes serielles Kabel!! |
35
|
Wed Sep 29 16:28:17 2010 |
Thomas Bergauer | common | EUDET telescope pedestal run | 20089 | 122850 (events built) | 29.09.2010 16:27 | 29.09.2010 17:33 |
EUDAQ config eudet2-test-6boards-in-2crates
rate: ~745.5 Hz
|
36
|
Wed Sep 29 17:38:15 2010 |
Thomas Bergauer | common | EUDET telescope pedestal run | 20097 | 118125 | 29.09.2010 17:37 | 29.09.2010 18:09 |
new pedestal run of EUDET telescope with new JTAG settings of plane 3 (sensor 11) [according to Ingrid]:
VRef2 98
VRef1A 87
VRef1B 143
VRef1C 108
VRef1D 153
|
38
|
Wed Sep 29 18:53:05 2010 |
Thomas Bergauer | Micron Baby | XY Positions | | | | |
center position: 113/47/7.5
top left (direction jura): 121/53.5/7.5
bottom right (direction saleve): 105/40.5/7.5 |
43
|
Thu Sep 30 14:45:16 2010 |
Thomas Bergauer | Micron Baby | run | run015 | 22.5k | 30.09.2010 13:18:11 | 30.09.2010 14:38:01 |
micron baby stack angle scan 15 grad mit 3 modulen
HV=100V, I=3.8uA
Beam on
telescope position: 129/76/187.5
telescope run 20109 |
48
|
Thu Sep 30 23:09:29 2010 |
Thomas Bergauer | Micron Baby | run | run019 | 162000 | 30.09.2010 23:08:03 | 01.10.2010 9:33:38 |
micron baby: 4 DUTs
HV=100V I=12uA I_stop=15.4uA
Beam on
45° rotation
telescope run 20115 (after JTAG programming of telescope chips due to config errors)
|
Attachment 1: run019_sensordisplay1.jpg
|
|
Attachment 2: Eudet-X_Cluster_Corr-run020115.jpg
|
|
Attachment 3: Eudet-Num_Clusters-run020115.jpg
|
|
53
|
Fri Oct 1 23:50:28 2010 |
Thomas Bergauer | CNM | run | run023 | 2275 | 01.10.2010 23:49:29 | 01.10.2010 23:53:12 |
CMN-SILC stack
HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA
too much optimized alignment (XY-table to 133/74/187.5) ;-)
telescope run 20119 |
54
|
Fri Oct 1 23:55:12 2010 |
Thomas Bergauer | CNM | run | run024 | 320k | 01.10.2010 23:53:12 | 02.10.2010 09:47:37 |
CMN-SILC stack
HV_SILC (Katze)=75V, 0.25uA (6 SiLC)
HV_CNM (Keithley)=30V, CNM_1=84nA, CNM_2=4.3uA
optimized alignment (XY-table to 133/76/187.5)
telescope run 20120 |
Attachment 1: cnm_2d_2.png
|
|
Attachment 2: cnm_2d_1.png
|
|