  HEPHY logbook for testbeam at SPS September/October 2010, Page 7 of 8  Not logged in ELOG logo
ID Date Author Project Subject Run Number Eventsup StartTime EndTime
  81   Mon Oct 4 20:16:19 2010 Christian IrmlerBELLEcalcal0118380004.10.2010 20:15:00 

angle scan Origami 1 and  Micron wedge 1
HV = 100V
chiller set to 14°C

HV connection of Origami 1 repaired

  87   Tue Oct 5 17:50:23 2010 Christian IrmlerITEcalcal0128380005.10.2010 17:50:2605.10.2010 18:09:48

6 x SILC + 2 x ITE
HV=100V
telescope position: 124/48

DUTs and SILC4/5 swapped
new module order (downstream): SILC0 - SILC3 - SILC1 - W7 - W9 - SILC2 - SILC4 -SILC5
SILC1 and SILC2 are horizontal

beam on

modules flushed with dry air (black tube)
rel. humidity: ~0% @ 26°C

  90   Wed Oct 6 11:03:59 2010 Marko DragicevicMicron Babycalcal0138380006.10.2010 11:03:12 

8 x Micron Baby (3 irradiated), no cooling
HV=100V I=115uA (start of run)
beam off

# VADJ/VPSP increased to 48 (was 30) for irradiated DUTs to restore baseline position (higher after irradiation)

ADC delay timing did not change after irradiation (verified)
 

  95   Wed Oct 6 23:03:15 2010 Marko DragicevicBELLEcalcal0148380006.10.2010 23:01:5606.10.2010 23:34:07

Belle stack without Origami2
HV=80V I=227uA

beam on

  97   Fri Oct 8 17:26:36 2010 Markus FriedlBELLEcalcal0158380008.10.2010 17:22:44 

Belle stack without Origami2
HV=11V I=32uA

beam on

2 suspicious APVs on n side of Micron due to burnt strip

trying to identify problematic Micron n side channels with low bias voltage (APV22 baseline saturates at higher bias)

  100   Fri Oct 8 23:05:03 2010 Markus FriedlBELLEcalcal0168380008.10.2010 23:03:16 

Belle stack w/o Origami2 after irradiation
HV=80V / 258 µA
cooling: Origami only, chiller set to -25°C,
2 suspicious APVs on n side of Micron due to burnt strip
 

Bias connection of Origami repaired --> lower noise on p-side

  104   Sat Oct 9 12:00:55 2010 Christian IrmlerBELLEcalcal0178380009.10.2010 11:59:37 

Origami1 (n-side only) + Micron1 (non-irrad)
angle scan: 73°
HV=100V / 40 µA
cooling: chiller set to 15°C

Origami p-side does not receive triggers -> disabled

origami flex straight part needed to be folded due to severe rotation angle -> nanonics connector problem...

  117   Mon Apr 11 18:25:35 2011 Manfred ValentanMicron BabyDUT5, source run, post irrad, pre zupfcal1048380011.04.2011 18:12:3811.04.2011 18:20:50

DUT5 (atoll)
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_single_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=26µA
I_end=48µA

 

  119   Wed Apr 13 12:13:09 2011 Manfred ValentanMicron BabyDUT5, source run, post irrad, post zupf ** DO NOT USE **cal1058380013.04.2011 12:02:4113.04.2011 12:10:55

DUT5 (atoll), each second bond removed to simulate intermediate strips
config file: D:/cern10/micronbaby/data_lab/DUT5/micronbaby_single_irrrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=18µA
I_end=21.5µA

data recorded with new switching power supplies and old intcal, which doesn't ignore bad strips!!!!!

  123   Tue May 17 13:02:34 2011 Manfred ValentanMicron BabyTRK1, source run, unirrad, prezupfcal1078380017.05.2011 12:2017.05.2011 12:35

TRK1: B2Micron_2826_1_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK1/prezupf/micronbaby_single.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.1µA
I_end=2µA

  125   Tue May 17 13:31:32 2011 Manfred ValentanMicron BabyTRK2, source run, unirrad, prezupfcal1088380017.05.2011 13:17:1717.05.2011 13:25:27

TRK2: B2Micron_2826_1_baby3 (atoll p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK2/prezupf/micronbaby_single.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.0µA
I_end=1.3µA

  127   Tue May 17 14:36:36 2011 Manfred ValentanMicron BabyTRK6, source run, unirrad, prezupfcal1098380017.05.2011 14:18:1317.05.2011 14:34:25

TRK6: B2Micron_2826_1_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK6/prezupf/micronbaby_multi6.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.3µA
I_end=1.8µA

  129   Tue May 17 15:10:52 2011 Manfred ValentanMicron BabyTRK7, source run, unirrad, prezupfcal1108380017.05.2011 15:16:2517.05.2011 15:24:35

TRK7: B2Micron_2826_19_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK7/prezupf/micronbaby_single.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.1µA
I_end=1.7µA

  131   Tue May 17 17:44:12 2011 Manfred ValentanMicron BabyGEW8, source run, unirrad, prezupfcal1118380017.05.2011 17:51:5017.05.2011 18:00:02

GEW8: B2Micron_2826_7_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/GEW8/prezupf/micronbaby_single.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.3µA
I_end=1.6µA

  133   Tue May 17 18:49:16 2011 Manfred ValentanMicron BabyB2Micron_2826_7_baby3, source run, unirrad, prezupfcal1128380017.05.2011 18:50:5017.05.2011 18:59:04

B2Micron_2826_7_baby3 (atoll p-stop)
config file: D:/cern10/micronbaby/data_lab/7baby3/prezupf/micronbaby_single.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.2µA
I_end=1.4µA

  136   Wed May 18 10:12:32 2011 Manfred ValentanMicron BabyDUT3, source run, unirrad, prezupfcal1138380018.05.2011 09:42:0818.05.2011 09:50:23

DUT3: B2Micron_2826_7_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/DUT3/prezupf/micronbaby_single_irrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=13µA
I_end=20µA

  138   Wed May 18 12:06:42 2011 Manfred ValentanMicron BabyDUT4, source run, irrad, prezupfcal1148380018.05.2011 11:42:2218.05.2011 11:50:36

DUT4: B2Micron_2826_19_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/DUT4/prezupf/micronbaby_single_irrad.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=50µA
I_end=104µA

  139   Wed May 18 14:40:33 2011 Manfred ValentanMicron BabyTRK6, source run, unirrad, postzupfcal1158380018.05.2011 13:41:4418.05.2011 13:49:58

TRK6: B2Micron_2826_1_baby4 (combined p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK6/postzupfzupf/micronbaby_single_postzupf.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.6µA
I_end=1.7µA

Every second bond zupfed. However, there was a mistake and the first ~60 channels are shifted by 1, see attachment!

Attachment 1: cal115_goodstrips_apv4.png
cal115_goodstrips_apv4.png
Attachment 2: cal115_goodstrips_apv5.png
cal115_goodstrips_apv5.png
  141   Wed May 18 15:36:03 2011 Manfred ValentanMicron BabyTRK7, source run, unirrad, postzupfcal1168380018.05.2011 14:43:1818.05.2011 14:51:31

TRK7: B2Micron_2826_19_baby2 (common p-stop)
config file: D:/cern10/micronbaby/data_lab/TRK7/postzupf/micronbaby_single_postzupf.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.6µA
I_end=1.7µA

every second bond zupfed...

Attachment 1: cal116_goodstrips_apv4.png
cal116_goodstrips_apv4.png
Attachment 2: cal116_goodstrips_apv5.png
cal116_goodstrips_apv5.png
  143   Wed May 18 16:31:12 2011 Manfred ValentanMicron BabyB2Micron_2826_7_baby3, source run, unirrad, prezupfcal1178380018.05.2011 15:44:5818.05.2011 15:53:10

B2Micron_2826_7_baby3 (atoll p-stop)
config file: D:/cern10/micronbaby/data_lab/7baby3/postzupf/micronbaby_single_postzupf.cfg

HV=+-50V
90Sr 1mCi ~centered
I_start=1.0µA
I_end=1.3µA

Every second bond zupfed...

Attachment 1: cal117_goodstrips_apv04.png
cal117_goodstrips_apv04.png
Attachment 2: cal117_goodstrips_apv05.png
cal117_goodstrips_apv05.png