Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-RD50-DESY-Apr2024 testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  Belle II and AIDA/Infineon Testbeam at DESY in Jan2014, Page 4 of 5  Not logged in ELOG logo
ID Date Author Project Subject Run Numberup Events StartTime EndTime Data
  24   Tue Jan 7 11:50:34 2014 Markus FriedlInfineonscanrun018x10k07.01.2014 11:xx:xx07.01.2014 11:xx:xxGood

Delay scan with single sample

Several runs

  25   Tue Jan 7 11:52:27 2014 Markus FriedlInfineonrunrun019650k07.01.2014 11:47:2407.01.2014 13:15:21Good

HV=300V  I=9.2uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single sample run

Beam energy = 3 GeV

  26   Tue Jan 7 13:20:48 2014 Markus FriedlInfineonrunrun020300k07.01.2014 13:18:5207.01.2014 14:35:49Good

HV=300V  I=8.5uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

multi6 with same settings as run019 (htr+3)

Beam energy = 3 GeV

  27   Tue Jan 7 14:47:38 2014 Markus FriedlInfineonrunrun021200k07.01.2014 14:46:0807.01.2014 15:13:23Good

HV=100V  I=7.2uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  29   Tue Jan 7 15:16:11 2014 Markus FriedlInfineonrunrun022200k07.01.2014 15:15:0907.01.2014 15:42:17Good

HV=150V  I=7.3uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  30   Tue Jan 7 15:54:22 2014 Markus FriedlInfineonrunrun023200k07.01.2014 15:53:0607.01.2014 16:25:26Good

HV=200V  I=8.0uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  31   Tue Jan 7 16:36:49 2014 Markus FriedlInfineonrunrun024200k07.01.2014 16:35:4407.01.2014 17:03:10Good

HV=210V  I=8.1uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  32   Tue Jan 7 17:09:56 2014 Markus FriedlInfineonrunrun025200k07.01.2014 17:08:5407.01.2014 17:36:12Good

HV=220V  I=8.1uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  33   Tue Jan 7 17:39:20 2014 Markus FriedlInfineonrunrun026200k07.01.2014 17:38:0107.01.2014 18:05:11Good

HV=230V  I=8.2uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  34   Tue Jan 7 18:11:52 2014 Markus FriedlInfineonrunrun027200k07.01.2014 18:10:4507.01.2014 18:37:42Good

HV=240V  I=8.2uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  35   Tue Jan 7 18:41:55 2014 Markus FriedlInfineonrunrun028200k07.01.2014 18:40:4007.01.2014 19:07:32Good

HV=250V  I=8.3uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  36   Tue Jan 7 19:12:07 2014 Markus FriedlInfineonrunrun029200k07.01.2014 19:11:0507.01.2014 19:38:00Good

HV=260V  I=8.7uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  37   Tue Jan 7 19:44:45 2014 Markus FriedlInfineonrunrun030200k07.01.2014 19:43:3507.01.2014 20:11:03Good

HV=270V  I=8.6uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  38   Tue Jan 7 20:26:50 2014 Markus FriedlInfineonrunrun031200k07.01.2014 20:25:3207.01.2014 20:58:47Good

HV=280V  I=8.4uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  39   Tue Jan 7 21:02:58 2014 Markus FriedlInfineonrunrun032200k07.01.2014 21:01:4207.01.2014 21:29:03Good

HV=290V  I=8.3uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

  40   Tue Jan 7 21:34:41 2014 Markus FriedlInfineonrunrun0334.117.024 (4M)07.01.2014 21:33:1208.01.2014 08:52:37Good

HV=300V  I=8.3uA

position adapted to center beam better (+10mm horizontal)

narrow trigger width (~5ns)

single trigger

 

Beam energy = 3 GeV

 

beam gone at 8:00 morning because of scheduled beam development. last few 100 events only noise.

  42   Wed Jan 8 14:10:07 2014 Thomas BergauerAIDArunrun03521168108.01.2014 14:01:2508.01.2014 17:03:55Unknown

 HV=100V  I=4.2uA

position adapted to center beam
 

accidentally hit reset button on PC which ended this run abruptly...

 

5 GeV beam

  43   Wed Jan 8 17:20:40 2014 Markus FriedlAIDArunrun036320k08.01.2014 17:18:5108.01.2014 22:14:58Unknown

 HV=100V  I=4.3uA

centered position

5GeV beam

narrow timing (trigger width), but 6 samples

 

  44   Wed Jan 8 22:19:11 2014 Markus FriedlAIDArunrun0371.2M08.01.2014 22:18:3009.01.2014 10:08:14Good

HV=100V  I=4.4uA
centered position
5GeV beam, single sample
narrow timing (trigger width)

 

  47   Thu Jan 9 11:04:34 2014 Markus FriedlAIDArunrun039100k09.01.2014 11:02:3609.01.2014 12:28:41Good

HV=10V  I=2.1uA
centered position, HV scan
5GeV beam, 6 samples
narrow timing (trigger width)

 

ELOG V3.1.5-fc6679b