  Alibava testbeam at DESY in Oct. 2017 (Dominic, Peter), Page 1 of 7  Not logged in ELOG logo
ID Date Author DUT Subject Run Number Eventsup XYPosition Cooling Data
  1   Thu Oct 12 10:06:31 2017 Dominic BloechcommonPacklist   NoUnknown

# Testbeam Packliste


### Alibava stuff

* Alibava Mainboard
* Alibava power supply 
* Alibava Modules: IFX_CenterBias, HPK_CenterBias, IFX_baby
* Flatband cable 2x
* Alibava resetbox + Cable
* USB AtoB 2x
* BIAS Cable for BeetleChip
* 2410 SMU + Power + USBtoRS232
* Alibava mounting stuff


### Tools and Stuff

* Labeling machine
* Duct tape
* 2 pairs of wool gloves
* Seitenschneider klein
* USBtoRS232 2x
* Lan cable
* Schiebelehre
* Imbuschlüssel
* Screwdriver set
* Special screwdriver set
* Cutter 2x
* labeling cartridge
* Lemo 16 ns
* Lemo 8 ns 2x
* Lemo 4 ns
* Lemo 2 ns

 

 

### PC and stuff

* DAQ PC + Cables

* Monitor + Cables
* Mouse
* Keyboard

 

### Data Storage
* Toshiba 1TB USB3
* Maxtor 4TB USB3

Attachment 1: ALiBaVa-System_klein.jpg
ALiBaVa-System_klein.jpg
Attachment 2: Stuff_klein.jpg
Stuff_klein.jpg
  2   Mon Oct 16 15:01:21 2017 Dominic BloechcommonPreparations   NoUnknown

### General

* Setup was set-up without problems. Pedestral and Calibration runs with ALIBAVA successfully.

** Warning: Trigger Out signal (busy signal) for the TLU is a TTL signal. Whereas the Trigger In signal can be a NIM signal. Either converting signal or adept in the TLU (we adepted the TLU matrix)

** TLU double triggered -> discriminated TLU signal beforehand 

*  Longterm BIAS Voltage for the IFX CenterBias to -100 V, slow control records IV-behavior in 10 sec steps.

* Security lectures taken

* 3 coincident measurements: 2 PM and the FEI4-Plane are connected to an "and" in the TLU

* Latency of Alibava without FEI4-Plane 133 [a.u] with FEI4 140 [a.u].

** Warning: This is due to the converter from the FEI4 which inverts the signal of the FEI4, this signal is then very noisy and would generate false positive triggers. -> Discriminate -> Long cables and latencies

### Other

* NI-Crate was broken (replaced by DESY)

Attachment 1: IMG_20171016_093005514.jpg
IMG_20171016_093005514.jpg
Attachment 2: IMG_20171016_150215807.jpg
IMG_20171016_150215807.jpg
Attachment 3: IMG_20171016_150230304.jpg
IMG_20171016_150230304.jpg
Attachment 4: IMG_20171016_150238370.jpg
IMG_20171016_150238370.jpg
  11   Wed Oct 18 12:28:08 2017 Peter PaulitschIFX_CenterBiasrun00010110110000122.9, 23.8, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000101.dat
Telescope filename  run000101.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -160 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 10000
Telescope triggers 14500
Data synced yes
Table position x/y 122.9, 23.8
Region of interest bias line
Trigger frequency  185 Hz
Meas. time 2 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 52.3%

 

  32   Fri Oct 20 14:57:30 2017 Peter PaulitschIFX_200_3e14run00018218210000075.93, 65.30, -115.55°YesGood
run settings
Parameter Value
Alibava filename  run000182.dat
Telescope filename  run000182.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.6%

 

  33   Fri Oct 20 14:58:56 2017 Peter PaulitschIFX_200_3e14run00018418410000075.93, 65.30, -110.55°YesGood
run settings
Parameter Value
Alibava filename  run000184.dat
Telescope filename  run000184.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.5%

 

  34   Fri Oct 20 15:00:11 2017 Peter PaulitschIFX_200_3e14run00018818810000075.93, 65.30, -100.55°YesGood
run settings
Parameter Value
Alibava filename  run000188.dat
Telescope filename  run000188.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

  35   Fri Oct 20 15:02:42 2017 Peter PaulitschIFX_200_3e14run00019019010000075.93, 65.30, -100.55°YesGood
run settings
Parameter Value
Alibava filename  run000190.dat
Telescope filename  run000190.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.9%

 

  36   Fri Oct 20 15:06:38 2017 Peter PaulitschIFX_200_3e14run00019319310000075.93, 65.30, -100.55°YesGood
run settings
Parameter Value
Alibava filename  run000193.dat
Telescope filename  run000193.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.3%

 

  40   Fri Oct 20 15:26:26 2017 Peter PaulitschIFX_200_3e14run00020120110000075.93, 65.30, -90.55°YesGood
run settings
Parameter Value
Alibava filename  run000201.dat
Telescope filename  run000201.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced no
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.7%

 

  53   Sat Oct 21 10:15:13 2017 Peter PaulitschHPK_3e14run00022822810000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000228.dat
Telescope filename  run000228.raw
Pedestal filename KIT_HPK_200_3e14.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  200
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 46.1%

 

  54   Sat Oct 21 10:17:03 2017 Peter PaulitschHPK_3e14run00023123110000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000231.dat
Telescope filename  run000231.raw
Pedestal filename KIT_HPK_200_3e14.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency  200
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 46.1%

 

  63   Mon Oct 23 12:19:57 2017 Peter PaulitschIFX_200_3e14run00018618810000075.93, 65.30, -100.55°YesGood
run settings
Parameter Value
Alibava filename  run000186.dat
Telescope filename  run000186.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.1%

 

  133   Mon Oct 30 12:04:14 2017 Peter PaulitschIFX_200_1e15run00038938910000076.50, 66.00, 0°YesUnknown
Parameter Value
Alibava filename  run000389.dat
Telescope filename  run000389.raw
Pedestal filename KIT_pedestal_600V_26102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature  
Latency 140 [a.u.]
Alibava triggers 100000
Telescope triggers 100000
Data synced yes
Table position x/y 76.50, 66.00
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency  
  41   Fri Oct 20 15:27:28 2017 Peter PaulitschIFX_200_3e14run000204204100000075.93, 65.30, -90.55°YesGood
run settings
Parameter Value
Alibava filename  run000204.dat
Telescope filename  run000204.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current 40 uA
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 1000000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 33.2%

 

  52   Sat Oct 21 10:14:07 2017 Peter PaulitschIFX_200_3e14run000225225100000075.93, 65.30, -70.55°YesGood
run settings
Parameter Value
Alibava filename  run000225.dat
Telescope filename  run000225.raw
Pedestal filename KIT_pedestal_600V_20102017.ped
Bias voltage -600 V
Dark current  
Sensor temperature -6
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 75.93, 65.30
Region of interest bias line
Trigger frequency   
Meas. time  
Beam energy 5.6 GeV
Cooling yes
Alibava efficiency 32.3%

 

  4   Tue Oct 17 15:13:42 2017 Dominic BloechIFX_CenterBiasrun000081811010000122.9, 26, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000081.dat
Telescope filename  run000081.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -162 nA
Sensor temperature 22 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  220 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.1%

 

  6   Wed Oct 18 11:48:51 2017 Peter PaulitschIFX_CenterBiasrun000085851010000122.9, 26, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000085.dat
Telescope filename  run000085.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.5 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.1%

 

  7   Wed Oct 18 11:59:30 2017 Peter PaulitschIFX_CenterBiasrun000088881010000122.9, 31, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000088.dat
Telescope filename  run000088.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.5 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 31
Region of interest upper part
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 45.4%

 

  8   Wed Oct 18 12:12:55 2017 Peter PaulitschIFX_CenterBiasrun000091911010000122.9, 21, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000091.dat
Telescope filename  run000091.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.6 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 21
Region of interest lower part
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.7%

 

  9   Wed Oct 18 12:19:14 2017 Peter PaulitschIFX_CenterBiasrun000093931010000122.9, 26, -3.3°NoGood
run settings
Parameter Value
Alibava filename  run000093.dat
Telescope filename  run000093.raw
Pedestal filename pedestal_17102017
Bias voltage -100 V
Dark current -165 nA
Sensor temperature 24.5 °C
Latency 140 [a.u.]
Alibava triggers 1010000
Telescope triggers 1000000
Data synced yes
Table position x/y 122.9, 26
Region of interest all
Trigger frequency  185 Hz
Meas. time 90 min
Beam energy 5.6 GeV
Cooling no
Alibava efficiency 44.1%

 