Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  Alibava/Mapsa testbeam at DESY in April 2017 (Axel, Grossi), Page 6 of 7  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Project Subjectup Run Number Events XYPosition Cooling Data
  138   Thu Apr 20 14:09:44 2017 Axel KoenigALiBaVa (IFX)run_011_Irrad_350V.dat01130000096, 64, 360NoUnknown

Remarks: Telescope run 641. Bias scan -350V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_011_Irrad_350V.dat
Telescope run 641
Triggers 300000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -350
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 37.9 uA
  86   Sun Apr 16 19:04:41 2017 Axel KoenigALiBaVa (IFX)run_012_Baby_60V.dat012135000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 589. Bias scan 60 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_012_Baby_60V.dat
Telescope run 589
Triggers 135000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -60
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  18   Thu Apr 13 16:00:25 2017 Axel KoenigALiBaVa (IFX)run_012_IFX_CB_70V.dat012175000109, 4.5, 240NoUnknown

Remarks: Telescope run 523. Bias scan 70 V.

Parameter Value
Alibava run run_012_IFX_CB_70V.dat
Telescope run 523
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -70
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  139   Thu Apr 20 15:08:33 2017 Axel KoenigALiBaVa (IFX)run_012_Irrad_250V.dat01230000096, 64, 360NoUnknown

Remarks: Telescope run 642. Bias scan -250V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_012_Irrad_250V.dat
Telescope run 642
Triggers 300000
Alibava trigger eff. ~66%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -250
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 32.1 uA
  87   Sun Apr 16 19:46:01 2017 Axel KoenigALiBaVa (IFX)run_013_Baby_65V.dat013112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 590. Bias scan 65 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_013_Baby_65V.dat
Telescope run 590
Triggers 112000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -65
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  21   Thu Apr 13 16:33:32 2017 Axel KoenigALiBaVa (IFX)run_013_IFX_CB_75V.dat013175000109, 4.5, 240NoUnknown

Remarks: Telescope run 524. Bias scan 75 V.

Parameter Value
Alibava run run_013_IFX_CB_75V.dat
Telescope run 524
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -75
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  140   Thu Apr 20 16:05:14 2017 Axel KoenigALiBaVa (IFX)run_013_Irrad_500V.dat01375000096, 54, 360NoUnknown

Remarks: Telescope run 643. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Positoj scan vertical.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_013_Irrad_500V.dat
Telescope run 643
Triggers 750000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 54 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
  88   Sun Apr 16 20:22:27 2017 Axel KoenigALiBaVa (IFX)run_014_Baby_70V.dat014250000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 591. Bias scan 70 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_014_Baby_70V.dat
Telescope run 591
Triggers 250000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -70
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  25   Thu Apr 13 17:11:31 2017 Axel KoenigALiBaVa (IFX)run_014_IFX_CB_80V.dat014175000109, 4.5, 240NoUnknown

3Remarks: Telescope run 525. Bias scan 80 V.

Parameter Value
Alibava run run_014_IFX_CB_80V.dat
Telescope run 525
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -80
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  142   Thu Apr 20 21:25:21 2017 Axel KoenigALiBaVa (IFX)run_014_Irrad_500V.dat01430000096, 69, 360NoUnknown

Remarks: Telescope run 646. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Positoj scan vertical.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_014_Irrad_500V.dat
Telescope run 646
Triggers 300000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 69 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
  90   Sun Apr 16 21:43:15 2017 Axel KoenigALiBaVa (IFX)run_015_Baby_100V.dat0152500000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 592. Long run, bias voltage 100 V. Long run finished without any problems.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_015_Baby_100V.dat
Telescope run 592
Triggers 2500000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  28   Thu Apr 13 17:47:28 2017 Axel KoenigALiBaVa (IFX)run_015_IFX_CB_90V.dat015175000109, 4.5, 240NoUnknown

3Remarks: Telescope run 526. Bias scan 90 V.

Parameter Value
Alibava run run_015_IFX_CB_90V.dat
Telescope run 526
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -90
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  141   Thu Apr 20 20:15:12 2017 Axel KoenigALiBaVa (IFX)run_015_Irrad_500V.dat01530000096, 64, 350NoUnknown

Remarks: Telescope run 647. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

rotation scan.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_015_Irrad_500V.dat
Telescope run 647
Triggers 300000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

350 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
  92   Mon Apr 17 08:14:24 2017 Axel KoenigALiBaVa (IFX)run_016_Baby_75V.dat016112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 593. Bias scan 75 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_016_Baby_75V.dat
Telescope run 593
Triggers 112000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -75
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  32   Thu Apr 13 18:26:57 2017 Axel KoenigALiBaVa (IFX)run_016_IFX_CB_100V.dat016750000109, 4.5, 240NoCrap

3Remarks: Telescope run 527-531. Bias scan 100 V. Telescope got out of sync. Match with telescope data only up to 175000 events.

Parameter Value
Alibava run run_016_IFX_CB_100V.dat
Telescope run 527-531
Triggers 750000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  143   Thu Apr 20 22:19:21 2017 Axel KoenigALiBaVa (IFX)run_016_Irrad_500V.dat016250000096, 64, 370NoUnknown

Remarks: Telescope run 648. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

rotation scan.

Telescope chiller turned on (15 deg).

Run finished over night!

Parameter Value
Alibava run run_016_Irrad_500V.dat
Telescope run 648
Triggers 2500000
Alibava trigger eff. ~70%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

370 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
  93   Mon Apr 17 09:01:09 2017 Axel KoenigALiBaVa (IFX)run_017_Baby_80V.dat017112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 594. Bias scan 80 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_017_Baby_80V.dat
Telescope run 594
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -80
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

  33   Thu Apr 13 20:49:30 2017 Axel KoenigALiBaVa (IFX)run_017_IFX_CB_110V.dat017175000109, 4.5, 240NoUnknown

Remarks: Telescope run 532. Bias scan 110 V.

Parameter Value
Alibava run run_017_IFX_CB_110V.dat
Telescope run 532
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -110
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  144   Fri Apr 21 07:46:57 2017 Axel KoenigALiBaVa (IFX)run_017_Irrad_500V.dat01720000096, 64, 360NoUnknown

Remarks: Telescope run 649. Bias voltage -500V, central position of sensor (DUT parallel to telescope planes at 360 deg due to different fixation compared to other DUTs). Telescope geometry identical to double irrad module.

Backside Biasing!

Beam energy scan.

Telescope chiller turned on (15 deg).

Parameter Value
Alibava run run_017_Irrad_500V.dat
Telescope run 649
Triggers 200000
Alibava trigger eff. ~70%
Beam energy 5.4 GeV
Telescope threshold 6
Bias voltage -500
Latency 132
Table x 96 mm
Table z 64 mm
Tabe rot

360 deg

Temperature at DUT -14.5 +- 1 deg. celsius
Current 39.2 uA
  95   Mon Apr 17 09:29:40 2017 Axel KoenigALiBaVa (IFX)run_018_Baby_90V.dat018112000102.5, 22.5, 240NoUnknown

Remarks: Telescope run 595. Bias scan 90 V.

Telescope chiller turned off (could be the root cause of runs crashing).

Parameter Value
Alibava run run_018_Baby_90V.dat
Telescope run 595
Triggers 112000
Alibava trigger eff. ~90%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -90
Latency 132
Table x 102.5 mm
Table z 22.5 mm
Tabe rot

240 deg

ELOG V3.1.5-fc6679b