Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  Alibava/Mapsa testbeam at DESY in April 2017 (Axel, Grossi), Page 1 of 7  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Projectup Subject Run Number Events XYPosition Cooling Data
  1   Wed Apr 12 21:06:44 2017 Axel KönigALiBaVa (IFX)run_001_IFX_CB_100V.dat001200000109, 4.5, 240NoUnknown

First run for alibava modules with Sensor IFX_CB.

Beam energy = 5.6 GeV

Telescope threshold = 6

Telescope data file = 496

Bias voltage = -100V

Telescope synchornisation: yes

Position table (is center of sensor):

x=109mm

y=4.5mm

phi=240 deg

 

Attachment 1: 11.png
11.png
Attachment 2: 46.png
46.png
  4   Wed Apr 12 22:34:18 2017 Axel KönigALiBaVa (IFX)run_002_IFX_CB_100V.dat002100000109, 0.5, 240NoUnknown

Remarks: Beam position on only one chip.

Parameter Value
Alibava run run_002_IFX_CB_100V.dat
Telescope run 498
triggers 100000
Alibava trigger eff. ?
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100 V
Latency 132
Table x 109 mm
Table z 0.5 mm
Table rot 240 deg (parallel to tele. planes)
  5   Wed Apr 12 22:41:08 2017 Axel KönigALiBaVa (IFX)run_003_IFX_CB_100V.dat003100000109, 8.5, 240NoUnknown

Remarks: Beam position on only one chip.

Parameter Value
Alibava run run_003_IFX_CB_100V.dat
Telescope run 500
Triggers 100000
Alibava trigger eff ?
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100 V
Latency 132
Table x 109 mm
table z 8.5 mm
table rot 240 deg (parallel to tele. planes)
  8   Thu Apr 13 09:23:20 2017 Axel KönigALiBaVa (IFX)run_005_IFX_CB_10V.dat005285000109, 4.5, 240NoUnknown

Remarks: Telescope run 508. Bias scan 10 V.

Parameter Value
Alibava run run_005_IFX_CB_10V.dat
Telescope run 508
Triggers 285000
Alibava trigger eff. ~36.5%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -10
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  9   Thu Apr 13 10:03:29 2017 Axel KönigALiBaVa (IFX)run_001_IFX_CB_100V.dat001200000109, 4.5, 240NoUnknown

Remarks: Telescope run 496. Bias voltage 100 V. First run at DESY TB.

Parameter Value
Alibava run run_001_IFX_CB_100V.dat
Telescope run 496
Triggers 200000
Alibava trigger eff. ?
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  10   Thu Apr 13 10:58:57 2017 Axel KoenigALiBaVa (IFX)run_006_IFX_CB_20V.dat006185000109, 4.5, 240NoUnknown

Remarks: Telescope run 510. Bias scan 20 V.

Parameter Value
Alibava run run_006_IFX_CB_10V.dat
Telescope run 510
Triggers 185000
Alibava trigger eff. ~55.6%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -20
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  11   Thu Apr 13 11:25:32 2017 Axel KoenigALiBaVa (IFX)run_007_IFX_CB_30V.dat007170000109, 4.5, 240NoUnknown

Remarks: Telescope run 513. Bias scan 30 V.

Parameter Value
Alibava run run_007_IFX_CB_30V.dat
Telescope run 513
Triggers 170000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -30
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  12   Thu Apr 13 12:00:40 2017 Axel KoenigALiBaVa (IFX)run_008_IFX_CB_40V.dat008175000109, 4.5, 240NoUnknown

Remarks: Telescope run 515. Bias scan 40 V.

Parameter Value
Alibava run run_008_IFX_CB_40V.dat
Telescope run 515
Triggers 175000
Alibava trigger eff. ~58%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -40
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  13   Thu Apr 13 12:33:20 2017 Axel KoenigALiBaVa (IFX)run_009_IFX_CB_50V.dat009175000109, 4.5, 240NoUnknown

Remarks: Telescope run 517. Bias scan 50 V.

Parameter Value
Alibava run run_009_IFX_CB_50V.dat
Telescope run 517
Triggers 175000
Alibava trigger eff. ~58%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -50
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  15   Thu Apr 13 13:07:39 2017 Axel KoenigALiBaVa (IFX)run_010_IFX_CB_60V.dat010175000109, 4.5, 240NoUnknown

Remarks: Telescope run 521. Bias scan 60 V.

Parameter Value
Alibava run run_010_IFX_CB_60V.dat
Telescope run 521
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -60
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  17   Thu Apr 13 15:26:22 2017 Axel KoenigALiBaVa (IFX)run_011_IFX_CB_65V.dat011175000109, 4.5, 240NoUnknown

Remarks: Telescope run 522. Bias scan 65 V.

Parameter Value
Alibava run run_011_IFX_CB_65V.dat
Telescope run 522
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -65
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  18   Thu Apr 13 16:00:25 2017 Axel KoenigALiBaVa (IFX)run_012_IFX_CB_70V.dat012175000109, 4.5, 240NoUnknown

Remarks: Telescope run 523. Bias scan 70 V.

Parameter Value
Alibava run run_012_IFX_CB_70V.dat
Telescope run 523
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -70
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  21   Thu Apr 13 16:33:32 2017 Axel KoenigALiBaVa (IFX)run_013_IFX_CB_75V.dat013175000109, 4.5, 240NoUnknown

Remarks: Telescope run 524. Bias scan 75 V.

Parameter Value
Alibava run run_013_IFX_CB_75V.dat
Telescope run 524
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -75
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  25   Thu Apr 13 17:11:31 2017 Axel KoenigALiBaVa (IFX)run_014_IFX_CB_80V.dat014175000109, 4.5, 240NoUnknown

3Remarks: Telescope run 525. Bias scan 80 V.

Parameter Value
Alibava run run_014_IFX_CB_80V.dat
Telescope run 525
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -80
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  28   Thu Apr 13 17:47:28 2017 Axel KoenigALiBaVa (IFX)run_015_IFX_CB_90V.dat015175000109, 4.5, 240NoUnknown

3Remarks: Telescope run 526. Bias scan 90 V.

Parameter Value
Alibava run run_015_IFX_CB_90V.dat
Telescope run 526
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -90
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  32   Thu Apr 13 18:26:57 2017 Axel KoenigALiBaVa (IFX)run_016_IFX_CB_100V.dat016750000109, 4.5, 240NoCrap

3Remarks: Telescope run 527-531. Bias scan 100 V. Telescope got out of sync. Match with telescope data only up to 175000 events.

Parameter Value
Alibava run run_016_IFX_CB_100V.dat
Telescope run 527-531
Triggers 750000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  33   Thu Apr 13 20:49:30 2017 Axel KoenigALiBaVa (IFX)run_017_IFX_CB_110V.dat017175000109, 4.5, 240NoUnknown

Remarks: Telescope run 532. Bias scan 110 V.

Parameter Value
Alibava run run_017_IFX_CB_110V.dat
Telescope run 532
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -110
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  36   Thu Apr 13 21:23:11 2017 Axel KoenigALiBaVa (IFX)run_018_IFX_CB_120V.dat018175000109, 4.5, 240NoUnknown

Remarks: Telescope run 533. Bias scan 120 V.

Parameter Value
Alibava run run_018_IFX_CB_120V.dat
Telescope run 533
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -120
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  59   Fri Apr 14 08:59:26 2017 Axel KoenigALiBaVa (IFX)run_020_IFX_CB_100V.dat020190000109, 4.5, 240NoCrap

Remarks: Telescope run 535. Bias scan 100 V. Run crashed over night. Telescope sync only up to event 190000 (planned 2500000 events).

Parameter Value
Alibava run run_020_IFX_CB_100V.dat
Telescope run 535
Triggers 190000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -100
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

  60   Fri Apr 14 09:25:09 2017 Axel KoenigALiBaVa (IFX)run_024_IFX_CB_140V.dat024175000109, 4.5, 240NoUnknown

3Remarks: Telescope run 547. Bias scan 140 V.

Parameter Value
Alibava run run_024_IFX_CB_140V.dat
Telescope run 547
Triggers 175000
Alibava trigger eff. ~60%
Beam energy 5.6 GeV
Telescope threshold 6
Bias voltage -140
Latency 132
Table x 109 mm
Table z 4.5 mm
Tabe rot

240 deg (parallel to tele. planes)

ELOG V3.1.5-fc6679b