Home CMS Production Clean room MedAustron HEPHY testbeams old
testbeam-RD50-DESY-Apr2024 testbeam-RD50-DESY-Jul2023 testbeam-RD50-CERN-Oct2022 testbeam-DESY-Oct2017 testbeam-DESY-April2017 testbeam-SPS2015 testbeam-SPS2014 testbeam-DESY14 testbeam-SPS12 testbeam-SPS11 testbeam-SPS10 testbeam-SPS09 testbeam-SPS08 LP-TPC
  Alibava/Mapsa testbeam at DESY in April 2017 (Axel, Grossi), Page 1 of 7  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
Entry  Thu Apr 13 18:26:57 2017, Axel Koenig, ALiBaVa (IFX), run_016_IFX_CB_100V.dat, 016, 750000, 109, 4.5, 240, No, Crap 
3Remarks: Telescope run 527-531. Bias scan 100 V. Telescope got out of sync. Match with telescope data only up to 175000 events.




Parameter
Value


Alibava run
run_016_IFX_CB_100V.dat


Entry  Fri Apr 14 08:59:26 2017, Axel Koenig, ALiBaVa (IFX), run_020_IFX_CB_100V.dat, 020, 190000, 109, 4.5, 240, No, Crap 
Remarks: Telescope run 535. Bias scan 100 V. Run crashed over night. Telescope sync only up to event 190000 (planned 2500000 events).




Parameter
Value


Alibava run
run_020_IFX_CB_100V.dat


Entry  Fri Apr 14 14:49:09 2017, Axel Koenig, ALiBaVa (IFX), run_023_IFX_CB_100V.dat, 023, 385000, 109, 0, 240, No, Crap 
Remarks: Telescope run 546. Bias voltage 100 V. Beam position on only one chip. Run destroyed. Intended trigger count 500k




Parameter
Value


Alibava run
run_023_IFX_CB_100V.dat


Entry  Fri Apr 14 21:11:18 2017, Axel Koenig, ALiBaVa (IFX), run_028_IFX_CB_100V.dat, 028, 3000000, 109, 4.5, 240, No, Crap 
Remarks: Telescope run 551. Longrun over night. Run crashed. Events only up to 2288000 valid.




Parameter
Value


Alibava run
run_028_IFX_CB_100V.dat


Entry  Sat Apr 15 21:18:26 2017, Axel Koenig, ALiBaVa (IFX), run_007_Baby_10V.dat, 007, 450000, 102.5, 22.5, 240, No, Crap 
Remarks: Telescope run 579. Bias scan 10 V. run crashed. Data vaild up to 274915.

Telescope chiller turned off (could be the root cause of runs crashing).




Parameter
Entry  Wed Apr 12 21:06:44 2017, Axel König, ALiBaVa (IFX), run_001_IFX_CB_100V.dat, 001, 200000, 109, 4.5, 240, No, Unknown 11.png46.png
First run for alibava modules with Sensor IFX_CB.

Beam energy = 5.6 GeV

Telescope threshold = 6
Entry  Wed Apr 12 21:55:13 2017, Axel König, common, pulse shape , , , , No, Unknown 21.png
pulse shape looks good for latency 132. This value is now taken by default for all runs unless otherwise stated.
Entry  Wed Apr 12 21:56:48 2017, Axel König, common, Kazu beetle settings, , , , No, Unknown 
Beetle setting for a better defined pulse shape. These parameters were found by kazu, a beetle developer at lhcb.

CompCtrl=9
Ibuf=10
Entry  Wed Apr 12 22:34:18 2017, Axel König, ALiBaVa (IFX), run_002_IFX_CB_100V.dat, 002, 100000, 109, 0.5, 240, No, Unknown 
Remarks: Beam position on only one chip.




Parameter
Value


Alibava run
Entry  Wed Apr 12 22:41:08 2017, Axel König, ALiBaVa (IFX), run_003_IFX_CB_100V.dat, 003, 100000, 109, 8.5, 240, No, Unknown 
Remarks: Beam position on only one chip.




Parameter
Value


Alibava run
run_003_IFX_CB_100V.dat


Entry  Thu Apr 13 09:23:20 2017, Axel König, ALiBaVa (IFX), run_005_IFX_CB_10V.dat, 005, 285000, 109, 4.5, 240, No, Unknown 
Remarks: Telescope run 508. Bias scan 10 V.




Parameter
Value


Alibava run
run_005_IFX_CB_10V.dat


Entry  Thu Apr 13 10:03:29 2017, Axel König, ALiBaVa (IFX), run_001_IFX_CB_100V.dat, 001, 200000, 109, 4.5, 240, No, Unknown 
Remarks: Telescope run 496. Bias voltage 100 V. First run at DESY TB.




Parameter
Value


Alibava run
run_001_IFX_CB_100V.dat


Entry  Thu Apr 13 10:58:57 2017, Axel Koenig, ALiBaVa (IFX), run_006_IFX_CB_20V.dat, 006, 185000, 109, 4.5, 240, No, Unknown 
Remarks: Telescope run 510. Bias scan 20 V.




Parameter
Value


Alibava run
run_006_IFX_CB_10V.dat


Entry  Thu Apr 13 11:25:32 2017, Axel Koenig, ALiBaVa (IFX), run_007_IFX_CB_30V.dat, 007, 170000, 109, 4.5, 240, No, Unknown 
Remarks: Telescope run 513. Bias scan 30 V.




Parameter
Value


Alibava run
run_007_IFX_CB_30V.dat


Entry  Thu Apr 13 12:00:40 2017, Axel Koenig, ALiBaVa (IFX), run_008_IFX_CB_40V.dat, 008, 175000, 109, 4.5, 240, No, Unknown 
Remarks: Telescope run 515. Bias scan 40 V.




Parameter
Value


Alibava run
run_008_IFX_CB_40V.dat


Entry  Thu Apr 13 12:33:20 2017, Axel Koenig, ALiBaVa (IFX), run_009_IFX_CB_50V.dat, 009, 175000, 109, 4.5, 240, No, Unknown 
Remarks: Telescope run 517. Bias scan 50 V.




Parameter
Value


Alibava run
run_009_IFX_CB_50V.dat


Entry  Thu Apr 13 13:07:17 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 1, 1000000, 0, No, Unknown 
Telescope 1209
Entry  Thu Apr 13 13:07:39 2017, Axel Koenig, ALiBaVa (IFX), run_010_IFX_CB_60V.dat, 010, 175000, 109, 4.5, 240, No, Unknown 
Remarks: Telescope run 521. Bias scan 60 V.




Parameter
Value


Alibava run
run_010_IFX_CB_60V.dat


Entry  Thu Apr 13 14:54:15 2017, Johannes Grossmann, MaPSA, Teleskope Correlation, 4, 1000000, 0, No, Unknown 
Telescope 1209

threshold 90 

 
Entry  Thu Apr 13 15:26:22 2017, Axel Koenig, ALiBaVa (IFX), run_011_IFX_CB_65V.dat, 011, 175000, 109, 4.5, 240, No, Unknown 
Remarks: Telescope run 522. Bias scan 65 V.




Parameter
Value


Alibava run
run_011_IFX_CB_65V.dat


ELOG V3.1.5-fc6679b