  18   Wed May 7 14:36:29 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 1 of 6 room temperature measurements using Sr90 Source (multi6)

Sr90 Radioactive Source
No cooling
Module position: -5:-20(top left)
HV Turned to +40,-40V with currents. +1.2 and -1.2 µA

N+1.25V current: 0.22A
N+2.5 V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A

Events: 200000

  19   Wed May 7 15:04:32 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 2 of 6 room temperature measurements using Sr90 Source (multi6)

Sr90 Radioactive Source
No cooling
Module position: -32:-20(top center)
HV Turned to +40,-40V with currents. +1.23 and -1.25 µA

N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.71A

Events: 200000

  20   Wed May 7 15:53:38 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 3 of 6 room temperature measurements using Sr90 Source (multi6)

Sr90 Radioactive Source
No cooling
Module position: -59:-20(top right)
HV Turned to +40,-40V with currents. +1.25 and -1.26 µA

N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.71A

Events: 200000

  21   Wed May 7 16:43:37 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 4 of 6 room temperature measurements using Sr90 Source (multi6)

Sr90 Radioactive Source
No cooling
Module position: -5:-30(bottom left)
HV Turned to +40,-40V with currents. +1.25 and -1.26 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A

Events: 200000

  22   Wed May 7 17:33:21 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 5 of 6 room temperature measurements using Sr90 Source (multi6)

Sr90 Radioactive Source
No cooling
Module position: -32:-30(bottom center)
HV Turned to +40,-40V with currents. +1.22 and -1.24 µA

N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A

Events: 200000

  23   Wed May 7 18:20:36 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 6 of 6 room temperature measurements using Sr90 Source (multi6)

Sr90 Radioactive Source
No cooling
Module position: -57:-30(bottom right)
HV Turned to +40,-40V with currents. +1.2 and -1.22 µA

N+1.25V current: 0.22A
N+2.50V current: 0.56A
P+1.25V current: 0.26A
P+2.50V current: 0.70A

Events: 200000

  26   Thu May 8 16:51:03 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 1 of 6 room temperature measurements using Sr90 Source (single)

Sr90 Radioactive Source
No cooling
Module position: 0:0(top left)
HV Turned to +40,-40V with currents. +1.0 and -1.0 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A

Events: 200000

  27   Thu May 8 16:51:57 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 2 of 6 room temperature measurements using Sr90 Source (single)

Failed with Error: "***ERROR*** Init NECO failed,  Return code: -32761"

Restarted and it works again

Sr90 Radioactive Source
No cooling
Module position: -27:0(top center)
HV Turned to +40,-40V with currents. +0.97 and -1.0 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.73A

Events: 200000

  28   Fri May 9 15:12:44 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 2 of 6 room temperature measurements using Sr90 Source (single)

Sr90 Radioactive Source
No cooling
Module position: -27:0(top center)
HV Turned to +40,-40V with currents. +0.95 and -0.98 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.72A

Events: 200000

  29   Fri May 9 16:44:41 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 3 of 6 room temperature measurements using Sr90 Source (single)

Sr90 Radioactive Source
No cooling
Module position: -54:0(top center)
HV Turned to +40,-40V with currents. +1.2 and -1.22 µA

N+1.25V current: 0.22A
N+2.50V current: 0.54A
P+1.25V current: 0.26A
P+2.50V current: 0.68A

Events: 200000

  30   Fri May 16 11:37:05 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 4 of 6 room temperature measurements using Sr90 Source (single)

Sr90 Radioactive Source
No cooling
Module position: 0:-10(bottom left)
HV Turned to +40,-40V with currents. +0.85 and -0-9 µA

N+1.25V current: 0.22A
N+2.50V current: 0.57A
P+1.25V current: 0.26A
P+2.50V current: 0.71A

Events: 200000

  31   Fri May 16 11:38:44 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 5 of 6 room temperature measurements using Sr90 Source (single)

Sr90 Radioactive Source
No cooling
Module position: -27:-10(bottom center)
HV Turned to +40,-40V with currents. +1.05 and -1.1 µA

N+1.25V current: 0.22A
N+2.50V current: 0.55A
P+1.25V current: 0.26A
P+2.50V current: 0.67A

Events: 200000

  32   Fri May 16 11:40:29 2014 Benedikt WürknerBelle IIsourceL3 moduleNo 6 of 6 room temperature measurements using Sr90 Source (single)

Sr90 Radioactive Source
No cooling
Module position: -54:-10(bottom right)
HV Turned to +40,-40V with currents. +1.1 and -1.1 µA

N+1.25V current: 0.22A
N+2.50V current: 0.55A
P+1.25V current: 0.26A
P+2.50V current: 0.67A

Events: 200000
 

  33   Thu Jun 5 10:33:46 2014 Benedikt WürknerBelle IIsourceSilc ModuleSilc Angle Measurement 1°

Measured the Silc 03/10 Module using the Sr90 Source to have a comparison for the Eta-Distribution at different angles. 

Data can be found on heros in: /home/medialib/LAB_Silc_Angle. 

Plots made with TuxOA for all different regions can be found in /home/users/bwuerkner/plots/. 

 

  34   Thu Jun 5 10:34:06 2014 Benedikt WürknerBelle IIsourceSilc ModuleSilc Angle Measurement 4°

Measured the Silc 03/10 Module using the Sr90 Source to have a comparison for the Eta-Distribution at different angles. 

Data can be found on heros in: /home/medialib/LAB_Silc_Angle. 

Plots made with TuxOA for all different regions can be found in /home/users/bwuerkner/plots/. 

 

  35   Thu Jun 5 10:34:29 2014 Benedikt WürknerBelle IIsourceSilc ModuleSilc Angle Measurement 7°

Measured the Silc 03/10 Module using the Sr90 Source to have a comparison for the Eta-Distribution at different angles. 

Data can be found on heros in: /home/medialib/LAB_Silc_Angle. 

Plots made with TuxOA for all different regions can be found in /home/users/bwuerkner/plots/. 

 

  36   Thu Jun 5 10:34:45 2014 Benedikt WürknerBelle IIsourceSilc ModuleSilc Angle Measurement 10°

Measured the Silc 03/10 Module using the Sr90 Source to have a comparison for the Eta-Distribution at different angles. 

Data can be found on heros in: /home/medialib/LAB_Silc_Angle. 

Plots made with TuxOA for all different regions can be found in /home/users/bwuerkner/plots/. 

 

  3   Wed Apr 23 13:05:05 2008 Christian IrmlerSPS Testbeam June08sourcehybrid 01time correlation between TDC and sensor measurement -> 3.1 ns RMS
  6   Wed May 7 15:24:49 2008 Christian IrmlerSPS Testbeam June08modulemodule 10/02properties (noise, intcal), APVs bonded to the sensor
Module tested with 1 and 2 rows bonded to the sensor, respectively. HV = 100 V Ibias (100V) = 19.1 nA Ibias (200V) = 23.7 nA
  9   Wed May 7 15:26:24 2008 Christian IrmlerSPS Testbeam June08modulemodule 20/09properties (noise, intcal), APVs bonded to the sensor
Module tested with 1 and 2 rows bonded to the sensor, respectively. HV = 100 V Ibias (100 V) = 25.1 nA Ibias (200 V) = 31.4 nA