Home CMS Production Clean room MedAustron HEPHY testbeams old
electronics module assembly SiDDaTA scratch
  HEPHY logbook of the Electronics Group  Not logged in ELOG logo
Message ID: 14     Entry time: Tue May 20 14:27:50 2008
Type: Measurement 
Author: Markus Friedl 
Project: BELLE Upgrade 
Subject: analysis results of source test 
Object ID: micron 
Measurement Type: source 
Run Number:  
*** NOTE: AFTER THIS MEASUREMENT WE REALIZED THAT BIASING WAS NOT DONE PROPERLY
          HENCE THE RESULTS BELOW ARE NOT RELIABLE 
          (in fact it is surprising that they are not worse) ***


Please find the results of the lab source test on the new Micron module here.
It is read out with 3 + 3 APV chips on either side.

Results table of the source measurement:
                      p-side     n-side
 Cluster signal [e]    18361      19434
 Strip noise [e]        1142       1193
 Avg cluster width      1.91       1.30
 Single strip SNR       16.1       16.3
 Cluster SNR            11.6       14.3
 Strip pitch [um]       50.0      153.5     

Apparently, the double metal capacitance is not so bad as expected, even though the Micron sensor does not use
the hourglass crossing scheme. Presumably the dielectric between metal 1 and 2 is rather thick (several um).
Strip noise is roughly the same on both p and n side, so the difference in Cluster SNR (*) only stems from the
unequal cluster width (which is a result of the different pitches).

Peak time precision vs SNR (last plot below) is worse compared to the values obtained with various HPK sensors
in the November 2007 beam test at KEK. However, this is a comparison of source and beam and thus might not be
significant. Let's see what we will get in the SPS beam test next week.

(*) Cluster SNR := sum(signal) / (strip_noise * sqrt(cluster_width) )
Attachment 1: vie_micron_run002_signal.gif  27 kB  | Hide | Hide all | Show all
vie_micron_run002_signal.gif
Attachment 2: vie_micron_run002_noise.gif  29 kB  | Hide | Hide all | Show all
vie_micron_run002_noise.gif
Attachment 3: vie_micron_run002_hit.gif  49 kB  | Show | Hide all | Show all
Attachment 4: vie_micron_run002_clw.gif  30 kB  | Hide | Hide all | Show all
vie_micron_run002_clw.gif
Attachment 5: vie_micron_p_run002_tres.gif  90 kB  | Hide | Hide all | Show all
vie_micron_p_run002_tres.gif
Attachment 6: vie_micron_n_run002_tres.gif  89 kB  | Hide | Hide all | Show all
vie_micron_n_run002_tres.gif
Attachment 7: vie_micron_p_run002_ressnr.gif  9 kB  | Hide | Hide all | Show all
vie_micron_p_run002_ressnr.gif
Attachment 8: vie_micron_n_run002_ressnr.gif  9 kB  | Hide | Hide all | Show all
vie_micron_n_run002_ressnr.gif
Attachment 9: trms_vs_snr_micron_src.gif  13 kB  | Hide | Hide all | Show all
trms_vs_snr_micron_src.gif
ELOG V3.1.5-fc6679b